Background: Ventricular arrhythmias (VAs) in the months following continuous-flow left ventricular assist devices (LVADs) implantation are common and associated with increased morbidity and mortality. We sought to evaluate for changes in implantable cardioverter defibrillator (ICD) performance following LVAD implantation.
Methods: We retrospectively reviewed the ICD interrogation reports and medical records of patients who received a continuous-flow LVAD at the University of Chicago Medical Center between January 2008 and December 2010. Patients who did not have an ICD prior to LVAD implantation were excluded.
Results: Forty-four patients were included. Median right ventricular lead sensing amplitude decreased from 10.0 (interquartile range 7.6-13.6) mV prior to LVAD to 7.9 (5.9-9.3) mV after LVAD implantation (P < 0.01). Significant reductions in sensing threshold were associated with smaller lead diameters, smaller pace/sense electrode surface areas, and Medtronic ICD leads (Medtronic Inc., Minneapolis, MN, USA). Lead impedance and capture threshold did not change significantly. ICD-related adverse events occurred in 30% of patients and led to ICD modification in 18% of patients.
Conclusions: Implantation of a continuous-flow LVAD is associated with changes in the performance of preexisting ICDs. ICD-related adverse events were encountered following LVAD implantation in this cohort, and at times resulted in invasive and noninvasive ICD system modification. Reductions in ICD sensing threshold after LVAD implantation may adversely affect ICD function. Formal ICD interrogation in addition to the regular follow-up testing is warranted post-LVAD.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/pace.12290 | DOI Listing |
Life (Basel)
January 2025
Internal Medicine Department, College of Medicine, King Faisal University, Al-Ahsa 31982, Saudi Arabia.
Background: Heart failure (HF) is a chronic condition that significantly affects morbidity and mortality. For patients with end-stage HF who are not candidates for heart transplantation, left ventricular assist devices (LVADs) provide mechanical circulatory support as a long-term solution, known as destination therapy (DT).
Objective: This meta-analysis aims to synthesize evidence on the survival rates, complications, and quality-of-life improvements associated with LVADs used as destination therapy in patients with end-stage HF.
J Card Fail
January 2025
Columbia University Irving Medical Center, New York, NY. Electronic address:
Background: The benefit of implantable cardioverter-defibrillators (ICD) and cardiovascular resynchronization therapy (CRT-D) in patients supported with a HeartMate 3 left ventricular assist device (LVAD) remains uncertain.
Methods: An analysis of the MOMENTUM 3 randomized clinical trial and the first 1000 patients in the Continued Access Protocol trial. Patients were divided into three groups based on the presence of ICD and/or CRT-D: No device (n=153, 11%), ICD only (n=699, 50.
Surg Pract Sci
June 2024
Baylor Scott and White, The Heart Hospital, 4708 Alliance Blvd, Suite 540, Plano, TX, United States.
Introduction: Although left ventricular assist device (LVAD) implantation is associated with improved survival in patients with end-stage heart failure, the impact of preoperative pulmonary function on short-term outcomes is unclear.
Methods: We conducted a retrospective review of all primary LVAD implants at a single institution. Common measures of preoperative pulmonary function were evaluated.
Front Surg
January 2025
Department of Cardiovascular Medicine, The Second Xiangya Hospital, Central South University, Changsha, Hunan, China.
We report a case of a patient with dilated cardiomyopathy who experienced recurrent ventricular tachycardia (VT) and multiple defibrillations following CRT-D implantation. Due to worsening cardiac function, the patient required surgical implantation of a left ventricular assist device (LVAD) as a bridge to heart transplantation. During the procedure, we used the Ensite three-dimensional mapping system to perform activation and substrate mapping of the VT targets, followed by endocardial and epicardial cryoballoon ablation of clinical VT.
View Article and Find Full Text PDFESC Heart Fail
January 2025
Division of Cardiology, Department of Medicine, Hospital of the University of Pennsylvania, Perelman School of Medicine, Philadelphia, Pennsylvania, USA.
Aims: Right ventricular (RV) failure (RVF) after left ventricular assist device (LVAD) implant is an important cause of morbidity and mortality. Modern, data-driven approaches for defining and predicting RVF have been under-utilized.
Methods: Two hundred thirty-two patients were identified with a mean age of 55 years; 40 (17%) were women, 132 were (59%) Caucasian and 74 (32%) were Black.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!