Effect of facial material softness and applied force on face mask dead volume, face mask seal, and inhaled corticosteroid delivery through an idealized infant replica.

J Aerosol Med Pulm Drug Deliv

1 Aerosol Research Laboratory of Alberta, Department of Mechanical Engineering, University of Alberta, Edmonton, Alberta, Canada , T6G 2G8.

Published: August 2014

Background: During the aerosol delivery device design and optimization process, in vitro lung dose (LD) measurements are often performed using soft face models, which may provide a more clinically relevant representation of face mask dead volume (MDV) and face mask seal (FMS) than hard face models. However, a comparison of MDV, FMS, and LD for hard and soft face models is lacking.

Methods: Metal, silicone, and polyurethane represented hard, soft, and very soft facial materials, respectively. MDV was measured using a water displacement technique. FMS was measured using a valved holding chamber (VHC) flow rate technique. The LD of beclomethasone dipropionate (BDP) delivered via a 100-μg Qvar® pressurized metered dose inhaler with AeroChamber Plus® Flow-Vu® VHC and Small Mask, defined as that which passes through the nasal airways of the idealized infant geometry, was measured using a bias tidal flow system with a filter. MDV, FMS, and LD were measured at 1.5 lb and 3.5 lb of applied force. A mathematical model was used to predict LD based on experimental measurements of MDV and FMS.

Results: Experimental BDP LD measurements for ABS, silicone, and polyurethane at 1.5 lb were 0.9 (0.6) μg, 2.4 (1.9) μg, and 19.3 (0.9) μg, respectively. At 3.5 lb, the respective LD was 10.0 (1.5) μg, 13.8 (1.4) μg, and 14.2 (0.9) μg. Parametric analysis with the mathematical model showed that differences in FMS between face models had a greater impact on LD than differences in MDV.

Conclusions: The use of soft face models resulted in higher LD than hard face models, with a greater difference at 1.5 lb than at 3.5 lb. A lack of a FMS led to decreased dose consistency; therefore, a sealant should be used when measuring LD with a hard ABS or soft silicone face model at 1.5 lb of applied force or less.

Download full-text PDF

Source
http://dx.doi.org/10.1089/jamp.2013.1087DOI Listing

Publication Analysis

Top Keywords

face models
24
face mask
16
applied force
12
soft face
12
face
11
mask dead
8
dead volume
8
mask seal
8
idealized infant
8
fms hard
8

Similar Publications

Objectives: Clinical breast examination (CBE) open the pathway to early detection and diagnosis of breast cancer. This study examined barriers to CBE uptake in seven sub-Saharan African (SSA) countries.

Methods: Data from the most current Demographic and Health Surveys of Burkina Faso, Cote d'Ivoire, Ghana, and Kenya Mozambique, Senegal and Tanzania was used.

View Article and Find Full Text PDF

Purpose: To compare a novel high-resolution optical coherence tomography (OCT) with improved axial resolution (High-Res OCT) with conventional spectral-domain OCT (SD-OCT) with regard to their capacity to characterize the disorganization of the retinal inner layers (DRIL) in diabetic maculopathy.

Methods: Diabetic patients underwent multimodal retinal imaging (SD-OCT, High-Res OCT, and color fundus photography). Best-corrected visual acuity and diabetes characteristics were recorded.

View Article and Find Full Text PDF

Purpose: Prolonged exposure to broadband light with a short-wavelength (blue) or long-wavelength (orange/red) bias is known to impact eye growth and refraction, but the mechanisms underlying this response are unknown. Thus, the present study investigated the effects of broadband blue and orange lights with well-differentiated spectrums on refractive development and global flash electroretinography (gfERG) measures of retinal function in the chick myopia model.

Methods: Chicks were raised for 4 days with monocular negative lenses, or no lens, under blue, orange, or white light.

View Article and Find Full Text PDF

Goals bias face perception.

J Exp Psychol Gen

January 2025

Department of Cognitive and Psychological Sciences, Brown University.

Faces-the most common and complex stimuli in our daily lives-contain multidimensional information used to infer social attributes that guide consequential behaviors, such as deciding who to trust. Decades of research illustrates that perceptual information from faces is processed holistically. An open question, however, is whether goals might impact this perceptual process, influencing the encoding and representation of the complex social information embedded in faces.

View Article and Find Full Text PDF

Volume electron microscopy (vEM) enables biologists to visualize nanoscale 3D ultrastructure of entire eukaryotic cells and tissues prepared by heavy atom staining and plastic embedding. The highest resolution vEM technique is focused ion-beam scanning electron microscopy (FIB-SEM), which provides nearly isotropic (~5-10 nm) spatial resolution at fluences of > 10,000 e /nm . However, it is not clear how such high resolution is achievable because serial block-face (SBF) SEM, which incorporates an in-situ ultramicrotome instead of a Ga FIB beam, results in radiation-induced collapse of similar specimen blocks at fluences of only ~20 e /nm .

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!