Purpose: Determination of dose error margins in radiation therapy planning due to variations in Hounsfield Units (HU) values dependent on the use of different CT scanning protocols.
Patients And Methods: Based on a series of different CT scanning protocols used in clinical practice, conversion tables for radiation dose calculations were generated and subsequently tested on a phantom. These tables were then used to recalculate the radiation therapy plans of 28 real patients after an incorrect scanning protocol had inadvertently been used for these patients.
Results: Different CT parameter settings resulted in errors of HU values of up to 2.6% for densities of <1.1 g/cm(3), but up to 25.6% for densities of >1.1 g/cm(3). The largest errors were associated with changes in the tube voltage. Tests on a virtual water phantom with layers of variable thickness and density revealed a sawtooth-shaped curve for the increase of dose differences from 0.3 to 0.6% and 1.5% at layer thicknesses of 1, 3, and 7 cm, respectively. Use of a beam hardening filter resulted in a reference dose difference of 0.6% in response to a density change of 5%. The recalculation of data from 28 patients who received radiation therapy to the head revealed an overdose of 1.3 ± 0.4% to the bone and 0.7 ± 0.1% to brain tissue. On average, therefore, one monitor unit (range 0-3 MU) per 100 MU more than the correct dose had been given.
Conclusion: Use of different CT scanning protocols leads to variations of up to 20% in the HU values. This can result in a mean systematic dose error of 1.5%. Specific conversion tables and automatic CT scanning protocol recognition could reduce dose errors of these types.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00066-013-0464-5 | DOI Listing |
A high-performance polarization stabilizer is highly desirable for many fields, such as optical communication, optical remote sensing, and optical measurement systems. It should simultaneously feature high tracking speed, compact size, low insertion loss, and low cost. Here, we experimentally demonstrate an integrated polarization stabilizer based on thin-film lithium niobate (TFLN) photonics and look-up table (LUT) method.
View Article and Find Full Text PDFAm J Physiol Endocrinol Metab
January 2025
Human and Animal Physiology, Wageningen University, Wageningen, The Netherlands.
The Respiratory Exchange Ratio (RER), which is the ratio of total carbon dioxide produced over total oxygen consumed, serves as a qualitative measure to determine the substrate usage of a particular organism on the whole-body level. Quantification of RER by its direct conversion into %Glucose (%G) and %Lipid oxidation (%L) at a given timepoint can be done by utilizing nonprotein respiratory quotient tables. These tables, however, are limited to specific increments, and intermediate RER values are not covered by these tables.
View Article and Find Full Text PDFFront Transplant
January 2025
Department of Surgical, Medical, Biomolecular Pathology and Intensive Care, University of Pisa, Pisa, Italy.
Background And Aims: There is growing interest in the environmental impact of surgical procedures, yet more information is needed specifically regarding liver transplantation. This study aims to quantify the total greenhouse gas emissions, or carbon footprint, associated with adult whole-size liver transplantation from donors after brain death, including the relevant back-table graft preparation.
Methods: The carbon footprint was calculated retrospectively using a bottom-up approach.
Environ Evid
January 2025
Department of Soil and Environment, Swedish University of Agricultural Sciences, Box 7014, 750 07, Uppsala, Sweden.
Background: To align with climate goals, greenhouse gas (GHG) emissions from agriculture must be reduced significantly. Cultivated peatlands are an important source of such emissions. One proposed measure is to convert arable fields on peatlands to grassland, as the Intergovernmental Panel on Climate Change (IPCC) default emission factors (EF) for organic soils are lower from grasslands.
View Article and Find Full Text PDFPlast Reconstr Surg Glob Open
January 2025
Department of Plastic and Reconstructive Surgery, The Ohio State University Wexner Medical Center, Columbus, OH.
Background: The first class of integrated plastic surgery residency applicants with pass/fail (P/F) step 1 assessments occurred during the 2023-2024 Match cycle. This study analyzes the results of a 2024 postmatch survey to program directors (PDs) regarding the impact of the conversion to P/F step 1 scoring and compares the results to the 2019 survey that attempted to predict how this conversion would influence decision-making when it was initially announced.
Methods: A 26-item survey evaluating PDs' perspectives on applicant characteristics and the application process following P/F step 1 scoring was distributed March 2024.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!