Purpose: To develop and evaluate a free-breathing chemical-shift-encoded (CSE) spoiled gradient-recalled echo (SPGR) technique for whole-heart water-fat imaging at 3 Tesla (T).

Methods: We developed a three-dimensional (3D) multi-echo SPGR pulse sequence with electrocardiographic gating and navigator echoes and evaluated its performance at 3T in healthy volunteers (N = 6) and patients (N = 20). CSE-SPGR, 3D SPGR, and 3D balanced-SSFP with chemical fat saturation were compared in six healthy subjects with images evaluated for overall image quality, level of residual artifacts, and quality of fat suppression. A similar scoring system was used for the patient datasets.

Results: Images of diagnostic quality were acquired in all but one subject. CSE-SPGR performed similarly to SPGR with fat saturation, although it provided a more uniform fat suppression over the whole field of view. Balanced-SSFP performed worse than SPGR-based methods. In patients, CSE-SPGR produced excellent fat suppression near metal. Overall image quality was either good (7/20) or excellent (12/20) in all but one patient. There were significant artifacts in 5/20 clinical cases.

Conclusion: CSE-SPGR is a promising technique for whole-heart water-fat imaging during free-breathing. The robust fat suppression in the water-only image could improve assessment of complex morphology at 3T and in the presence of off-resonance, with additional information contained in the fat-only image.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4007374PMC
http://dx.doi.org/10.1002/mrm.24982DOI Listing

Publication Analysis

Top Keywords

fat suppression
16
technique whole-heart
8
whole-heart water-fat
8
water-fat imaging
8
patients cse-spgr
8
fat saturation
8
image quality
8
fat
6
whole-heart chemical
4
chemical shift
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!