Hb H Interference on Measurement Of HbA1c With Ion-Exchange HPLC.

Acta Inform Med

Department of Medical Biochemistry, Gulhane Military Medical Academy, Ankara, Turkey.

Published: October 2013

In this article, an interference caused by hemoglobin H (Hb H), during the measurement of hemoglobin A1c (HbA1c) with ion exchange high pressure liquid chromatography (HPLC) method, was presented in blood sample of a 20-year-old male patient. HbA1c measurement was performed with Agilent 1200 HPLC system using a commercial Recipe HbA1c ion-exchange column. Hemoglobin electrophoresis was performed with Interlab G26 agarose electrophoresis automated compact system. HbA1c level was 18.2% and HbA0 level was 81.5% with ion-exchange HPLC method. Patient's fasting serum glucose was assessed before HbA1c measurement and the result was 165 mg/dL (9.16 mmol/L). On the other hand, the result of HbA0 was 87.9%, Hb H was 10.8% and Hb A2 was 1.3% with electrophoresis. Whole blood test values were within reference ranges except MCV. MCV value was 79.6 fL. It is important to keep in mind that HbA1c level might be considered falsely high with ion-exchange HPLC method because of Hb H containing sample.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3804497PMC
http://dx.doi.org/10.5455/aim.2013.21.216-218DOI Listing

Publication Analysis

Top Keywords

ion-exchange hplc
12
hplc method
12
hba1c ion-exchange
8
hba1c measurement
8
hba1c level
8
hba1c
7
hplc
5
interference measurement
4
measurement hba1c
4
ion-exchange
4

Similar Publications

Bacteriocins, naturally derived antimicrobial peptides, are considered promising alternatives to traditional preservatives and antibiotics, particularly in food and medical applications. Despite extensive research on various bacteriocins, cyclic varieties remain understudied. This study introduces Gassericin GA-3.

View Article and Find Full Text PDF

Phytate in plants (inositol phosphates, InsPs) affects mineral bioavailability. However, methods for their quantification may lead to variable results, and some are nonspecific (spectrophotometric techniques). In this study, ion-pair high-performance liquid chromatography (HPLC) was coupled with post-column derivatization to allow fluorescence detection (FLD, λ324/λ364 nm) of InsPs.

View Article and Find Full Text PDF

Objectives: To describe a variant hemoglobin that interferes with HbA analysis by cation exchange HPLC.

Case Presentation: A 78 years-old Spanish male patient visited the Internal Medicine Clinic for a routine check-up, with HbA included to screen for diabetes. He had suffered hypertension and dyslipidemia, and the patient had no previous symptoms suggestive of diabetes such as hyperglycemia, weight loss, polydipsia, polyuria or tiredness.

View Article and Find Full Text PDF

A pectinase-producing strain, sp. X-8, was isolated from the soil for the first time. Subsequently, sp.

View Article and Find Full Text PDF

Hemoglobin disorders are highly prevalent inherited hematological defects in Southern China. The identification of novel variants in globin genes and accurate assessment of hematological parameters play a crucial role in precise genetic counseling and clinical practice. Peripheral blood samples were collected for hematological analysis, including red blood cell and hemoglobin assessment, while serum ferritin levels were measured to detect iron depletion.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!