Purpose: To assess the accuracy of dual-energy CT (DECT) for the quantification of iodine concentrations in a thoracic phantom across various cardiac DECT protocols and simulated patient sizes.

Materials And Methods: Experiments were performed on first- and second-generation dual-source CT (DSCT) systems in DECT mode using various cardiac DECT protocols. An anthropomorphic thoracic phantom was equipped with tubular inserts containing known iodine concentrations (0-20 mg/mL) in the cardiac chamber and up to two fat-equivalent rings to simulate different patient sizes. DECT-derived iodine concentrations were measured using dedicated software and compared to true concentrations. General linear regression models were used to identify predictors of measurement accuracy

Results: Correlation between measured and true iodine concentrations (n = 72) across CT systems and protocols was excellent (R = 0.994-0.997, P < 0.0001). Mean measurement errors were 3.0 ± 7.0% and -2.9 ± 3.8% for first- and second-generation DSCT, respectively. This error increased with simulated patient size. The second-generation DSCT showed the most stable measurements across a wide range of iodine concentrations and simulated patient sizes.

Conclusion: Overall, DECT provides accurate measurements of iodine concentrations across cardiac CT protocols, strengthening the case for DECT-derived blood volume estimates as a surrogate of myocardial blood supply.

Key Points: • Dual-energy CT provides new opportunities for quantitative assessment in cardiac imaging. • DECT can quantify myocardial iodine as a surrogate for myocardial perfusion. • DECT measurements of iodine concentrations are overall very accurate. • The accuracy of such measurements decreases as patient size increases.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4126413PMC
http://dx.doi.org/10.1007/s00330-013-3040-6DOI Listing

Publication Analysis

Top Keywords

iodine concentrations
16
accuracy dual-energy
8
thoracic phantom
8
cardiac dect
8
dect protocols
8
iodine
5
concentrations
5
dual-energy computed
4
computed tomography
4
tomography measurement
4

Similar Publications

Evaluation of charge summing correction in CdTe-based photon-counting detectors for breast CT: performance metrics and image quality.

J Med Imaging (Bellingham)

January 2025

U.S. Food and Drug Administration, Office of Science and Engineering Labs, Division of Imaging, Diagnostics, and Software Reliability, Silver Spring, Maryland, United States.

Purpose: We evaluate the impact of charge summing correction on a cadmium telluride (CdTe)-based photon-counting detector in breast computed tomography (CT).

Approach: We employ a custom-built laboratory benchtop system using the X-THOR FX30 0.75-mm CdTe detector (Varex Imaging, Salt Lake City, Utah, United States) with a pixel pitch of 0.

View Article and Find Full Text PDF

: We conducted a cross-sectional study to investigate whether children receiving long-term parenteral nutrition (LPN) are at risk of imbalances in selected trace elements. : Serum levels of manganese, zinc, copper, selenium, and iodine were measured in 83 children on LPN and compared with 121 healthy controls. Children with signs of infection or elevated C-reactive protein levels were excluded.

View Article and Find Full Text PDF

Iodine is a key micronutrient essential for the synthesis of thyroid hormone, which regulates metabolic processes and maintains overall health. Despite its importance, iodine deficiency is a global health issue, leading to disorders such as goiter, hypothyroidism, and developmental abnormalities. Biofortification of crops with iodine is a promising strategy to enhance the dietary iodine intake, providing an alternative to iodized salt.

View Article and Find Full Text PDF

Background/objectives: rapidly acquires antibiotic resistance and demonstrates increasing tolerance to antiseptics. This study evaluated the activity of eight antiseptics against , assessed its ability to develop adaptation to these antiseptics, and, for the first time, determined the Karpinski Adaptation Index (KAI) for this bacterium.

Methods: The minimal inhibitory concentration (MIC), susceptibility to antibiotics, bactericidal time according to EN 1040:2005, adaptation potential, and KAI of strains were evaluated.

View Article and Find Full Text PDF

A simple method for determining elemental sulfur in environmental water was developed and applied to seawater samples collected immediately after the occurrence of blue tides in Tokyo Bay. To investigate the concentration and extraction methods, artificial elemental sulfur was quantitatively produced by oxidizing a sulfide solution with an iodine solution, then used as a standard reagent in the experiments. To concentrate the elemental sulfur in the water sample, glass filter paper (GF/F) was used to filter and collect the elemental sulfur.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!