Objective: To compare the effectiveness, safety and cost of Tristel Fuse (chlorine dioxide) with Cidex OPA (ortho-phthaldehyde; 1,2-benzenedicarboxaldehyde) in an automated endoscopic reprocessor (AER) for high-level disinfection of flexible cystoscopes.

Patients And Methods: A randomised single-blind study comparing the high-level disinfectants Tristel Fuse as a simple office-based soak and Cidex OPA using an AER was performed. Participants were 'blinded' to the agent used for disinfection of the flexible cystoscopes. All patients had negative mid-stream urine at baseline, (MSU) no symptoms suggestive of urinary tract infection (UTI) on the day of investigation, no recent antibiotic use or current indwelling urinary catheter. Patients who underwent cystoscopic biopsy during the procedure were excluded. A urine analysis was done before and 3-5 days after cystoscopy and multiple equipment cultures were performed. The Urogenital Distress Inventory (UDI-6 + two questions from the 'long-form'), symptom and quality-of-life scores were assessed before and after cystoscopy as were ease-of-use assessments and a full cost analysis.

Results: In all, 180 of 465 screened participants were randomised 1:1 and the mean age was 72.1 years, 17% were females and 57% of procedures were performed for bladder tumour surveillance. The urine analysis was positive in 5.4% of patients in each group and 29% (Tristel) vs 20% (Cidex) of patients had urinary leukocyturia (p = ns) after cystoscopy. The turnover (minutes per cycle) was 7.5 (Tristel) vs 26.7 (Cidex). The per-procedure costs were $11.67 (American dollars) for Tristel Fuse and $21.82 for Cidex OPA with fixed costs of $4788 for Tristel Fuse and $60,514 for Cidex OPA.

Conclusions: Tristel Fuse appears to be as effective and more cost-effective than Cidex OPA for high-level disinfection of flexible cystoscopes. This has significant cost implications for the office urologist.

Download full-text PDF

Source
http://dx.doi.org/10.1111/bju.12208DOI Listing

Publication Analysis

Top Keywords

tristel fuse
24
cidex opa
20
disinfection flexible
16
high-level disinfection
12
flexible cystoscopes
12
randomised single-blind
8
tristel
8
fuse chlorine
8
chlorine dioxide
8
cidex
8

Similar Publications

Objective: To compare the effectiveness, safety and cost of Tristel Fuse (chlorine dioxide) with Cidex OPA (ortho-phthaldehyde; 1,2-benzenedicarboxaldehyde) in an automated endoscopic reprocessor (AER) for high-level disinfection of flexible cystoscopes.

Patients And Methods: A randomised single-blind study comparing the high-level disinfectants Tristel Fuse as a simple office-based soak and Cidex OPA using an AER was performed. Participants were 'blinded' to the agent used for disinfection of the flexible cystoscopes.

View Article and Find Full Text PDF

Development of a Ct equation for the inactivation of Cryptosporidium oocysts with chlorine dioxide.

Water Res

June 2003

Environmental Engineering and Public Health Consultant, 9627 Lansford Drive, Cincinnati, OH 45242, USA.

Cryptosporidium parvum, a protozoan parasite, has been implicated in a number of waterborne disease outbreaks and is difficult to inactivate using free chlorine. It appears, however, to be inactivated more easily by other oxidants such as chlorine dioxide or ozone. A major element of the EPA (US EPA) strategy for controlling C.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!