Scatter correction associated with dedicated dual-source CT hardware improves accuracy of lung air measures.

Acad Radiol

Department of Radiology, Division of Physiological Imaging, University of Iowa Hospitals and Clinics, 200 Hawkins Drive, CC 701 GH, University of Iowa Carver College of Medicine, Iowa City, IA 52241; Department of Biomedical Engineering, University of Iowa, Iowa City, IA.

Published: November 2013

Rationale And Objectives: Accurate assessment of air density used to quantitatively characterize amount and distribution of emphysema in chronic obstructive pulmonary disease (COPD) subjects has remained challenging. Hounsfield units (HU) within tracheal air can be considerably less negative than -1000 HU. This study has sought to characterize the effects of improved scatter correction used in dual-source pulmonary computed tomography (CT).

Materials And Methods: Dual-source dual-energy (DSDE) and single-source (SS) scans taken at multiple energy levels and scan settings were acquired for quantitative comparison using anesthetized ovine (n = 6), swine (n = 13), and a lung phantom. Data were evaluated for the lung, inferior vena cava, and tracheal segments. To minimize the effect of cross-scatter, the phantom scans in the DSDE mode were obtained by reducing the current of one of the tubes to near zero.

Results: A significant shift in mean HU values in the tracheal regions of animals and the phantom is observed, with values consistently closer to -1000 HU in DSDE mode. HU values associated with SS mode demonstrated a positive shift of up to 32 HU. In vivo tracheal air measurements demonstrated considerable variability with SS scanning, whereas these values were more consistent with DSDE imaging. Scatter effects in the lung parenchyma differed from adjacent tracheal measures.

Conclusion: Data suggest that the scatter correction introduced into the dual-energy mode of imaging has served to provide more accurate CT lung density measures sought to quantitatively assess the presence and distribution of emphysema in COPD subjects. Data further suggest that CT images, acquired without adequate scatter correction, cannot be corrected by linear algorithms given the variability in tracheal air HU values and the independent scatter effects on lung parenchyma.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3825679PMC
http://dx.doi.org/10.1016/j.acra.2013.04.018DOI Listing

Publication Analysis

Top Keywords

scatter correction
16
tracheal air
12
distribution emphysema
8
copd subjects
8
dsde mode
8
scatter effects
8
effects lung
8
lung parenchyma
8
scatter
6
tracheal
6

Similar Publications

The marginal wells in low-permeability oil fields are characterized by small storage size, scattered distribution, intermittent production, etc. The construction of large-scale gathering pipelines has large investment. So the current production mode is featured by single well tank oil storage, oil tank truck transportation and manual tank truck scheduling.

View Article and Find Full Text PDF

Linear digital filters are at the core of image reconstruction and processing for many coherent optical imaging techniques, such as digital holography (DH) or optical coherence tomography (OCT). They can also be efficiently implemented using fast Fourier transform (FFT) with appropriate transfer/filter functions that operate in the frequency domain. However, even with optimal filter design, they suffer from side effects such as sidelobe generation or resolution limitations, e.

View Article and Find Full Text PDF

Tetramethylammonium (TMA) is a ubiquitous cationic motif in biochemistry, found in the charged choline headgroup of membrane phospholipids and in tri-methylated lysine residues, which modulates histone-DNA interactions and impacts epigenetic mechanisms. TMA interactions with anionic species, particularly carboxylate groups of amino acid residues and extracellular sugars, are of substantial biological relevance, as these interactions mediate a wide range of cellular processes. This study investigates the molecular interactions between TMA and acetate, representing carboxylate-containing groups, using neutron scattering experiments complemented by force fields and molecular dynamics (MD) simulations.

View Article and Find Full Text PDF

Analyzing manure nutrients such as total ammonium nitrogen (NH), dry matter (DM), calcium oxide (CaO), total nitrogen (-N), phosphorus pentoxide (PO), magnesium oxide (MgO), and potassium oxide (KO) helps in fulfilling crop nutritional needs while improving the profitability and a lower risk of pollutants. This study used two Near Infra Red (NIR) spectral datasets of fresh and dried manure. The freshly prepared NHCl, CaO, Ca(OH), PO, MgO, and KO samples were used for spectral signature peak identification and calibration.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!