Purpose: Fractures of the femur associated with total hip arthroplasty are a significant concern in orthopaedic and trauma surgery. However, little is known about the different biomechanical properties of internal fixation systems in combination with periprosthetic fractures. In this study two new internal fixation systems for periprosthetic fractures are investigated using a cadaver fracture model simulating a Vancouver B1 periprosthetic femur fracture.

Methods: Nine pairs of fresh-frozen cadaver femurs were scanned by dual X-ray absorptiometry. Cementless total hip prostheses were implanted and a periprosthetic femur fracture was simulated. Fractures were randomly fixed either with the fixed angle locking attachment plate (LAP®, Depuy Synthes®, Solothurn, Switzerland) or the variable angle non-contact bridging plate (NCB®, Zimmer GmbH, Winterthur, Switzerland). Each construct was cyclically loaded to failure in axial compression.

Results: Axial stiffness and cycles to failure were significantly higher in the NCB group. Both systems were able to be fixed well around the femoral stem.

Conclusion: The two different internal fixation systems for periprosthetic fractures differed significantly in our setup. The non-contact bridging plate system revealed significantly higher failure load and may be the preferred option where high stability and load capacity is needed right after operation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3890125PMC
http://dx.doi.org/10.1007/s00264-013-2113-0DOI Listing

Publication Analysis

Top Keywords

periprosthetic femur
12
internal fixation
12
fixation systems
12
periprosthetic fractures
12
femur fracture
8
total hip
8
systems periprosthetic
8
non-contact bridging
8
bridging plate
8
periprosthetic
6

Similar Publications

Periprosthetic fractures and their associated complications present significant challenges for orthopaedic surgeons. It is important to provide an overview of the current management of periprosthetic fractures, including techniques for osteosynthesis and revision total hip and knee arthroplasty, as well as special considerations for periprosthetic acetabular fractures, periprosthetic tibial fractures, and interprosthetic femur fractures. In addition, the guiding principles for the management of potential subsequent complications including infection, nonunion, and instability are discussed.

View Article and Find Full Text PDF

Periprosthetic femur fractures around the hip are one of the most common aseptic complications following total hip arthroplasty. Understanding the risk factors of periprosthetic femur fracture can aid surgeons in the prevention of these injuries. The Vancouver classification provides a reproducible description of the factors that should be considered in the treatment of patients with periprosthetic femur fractures: fracture site, implant stability, and bone stock.

View Article and Find Full Text PDF

Background: When patients with a failed hip arthroplasty are unsuitable for reimplantation, Girdlestone resection arthroplasty (GRA) is a viable treatment option. We report on a patient who was treated with a GRA due to a periprosthetic infection. We discovered partial paralysis of the quadriceps muscle in this patient.

View Article and Find Full Text PDF

Periprosthetic joint infection (PJI) is a leading cause and major complication of joint replacement failure. As opposed to standard-of-care systemic antibiotic prophylaxis for PJI, we developed and tested titanium femoral intramedullary implants with titania nanotubes (TNTs) coated with the antibiotic gentamicin and slow-release agent chitosan through electrophoretic deposition (EPD) in a mouse model of PJI. We hypothesized that these implants would enable local gentamicin delivery to the implant surface and surgical site, effectively preventing bacterial colonization.

View Article and Find Full Text PDF

Total femoral replacement (TFR) is not a common surgery and most indications are for oncological pathologies. However, there are few instances where non-oncological indications might necessitate TFR; this may be a salvage surgery for failed previous hip and/or knee surgeries with consequent significant femur bone loss. We present a 59-year-old obese woman with right thigh pain and difficulty with walking of 5 years duration.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!