Electroencephalography (EEG) testing in clinical labs makes use of large amplifiers and complex software for data acquisition. While there are new ambulatory electroencephalogram (EEG) systems, few have been directly compared to a gold standard system. Here, an ultra-low power wireless EEG system designed by Imec is tested against the gold standard Neuroscan SynAmps2 EEG system, recording simultaneously from the same laboratory cap prepared with electrode gel. The data was analyzed using correlation analysis for both time domain and frequency domain data. The analysis indicated a high Pearson's correlation coefficient (mean=0.957, median=0.985) with high confidence (mean P=0.002) for 10-second sets of data transformed to the frequency domain. The time domain results had acceptable Pearson's coefficient (mean=0.580, median =0.706) with high confidence (mean P=0.008).