Objective: To evaluate the effect of continuous glucose monitoring (CGM) on the frequency of severe hypoglycemia (SH) in patients with established hypoglycemia unawareness.

Research Design And Methods: We conducted a retrospective audit of 35 patients with type 1 diabetes and problematic hypoglycemia unawareness, despite optimized medical therapy (continuous subcutaneous insulin infusion/multiple daily insulin injections), who used CGM for >1 year.

Results: Over a 1-year follow-up period, the median rates of SH were reduced from 4.0 (interquartile range [IQR] 0.75-7.25) episodes/patient-year to 0.0 (0.0-1.25) episodes/patient-year (P < 0.001), and the mean (±SD) rates were reduced from 8.1 ± 13 to 0.6 ± 1.2 episodes/year (P = 0.005). HbA1c was reduced from 8.1 ± 1.2% to 7.6 ± 1.0% over the year (P = 0.005). The mean Gold score, measured in 19 patients, did not change: 5.1 ± 1.5 vs. 5.2 ± 1.9 (P = 0.67).

Conclusions: In a specialist experienced insulin pump center, in carefully selected patients, CGM reduced SH while improving HbA1c but failed to restore hypoglycemia awareness.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3836107PMC
http://dx.doi.org/10.2337/dc13-0939DOI Listing

Publication Analysis

Top Keywords

continuous glucose
8
glucose monitoring
8
severe hypoglycemia
8
patients type
8
type diabetes
8
rates reduced
8
hypoglycemia
5
patients
5
real-time continuous
4
monitoring reduces
4

Similar Publications

This study evaluated a next-generation automated insulin delivery (AID) algorithm for Omnipod in type 1 and type 2 diabetes across multiple phases: 14-day run-in with usual therapy, 48-h AID use in a hotel setting (type 1 only), and up to 6 weeks of outpatient AID use. Participants did, or did not, deliver manual boluses at alternating periods. Twelve adults with type 1 diabetes completed the hotel phase; 9 of those 12 plus 8 adults with type 2 diabetes completed the subsequent outpatient phase.

View Article and Find Full Text PDF

The implanted Eversense Continuous Glucose Monitoring (CGM) System transitioned from 90- to 180- to 365-day durations marketed today. This report summarizes the 365-day clinical study. ENHANCE was a prospective, multicenter study evaluating the accuracy and safety of the Eversense 365 CGM system through 1 year in adults with diabetes.

View Article and Find Full Text PDF

Impact of time in tight range on all-cause and cardiovascular mortality in type 2 diabetes: A prospective cohort study.

Diabetes Obes Metab

January 2025

Department of Endocrinology and Metabolism, Shanghai Sixth People's Hospital Affiliated to Shanghai Jiao Tong University School of Medicine, Shanghai Clinical Center for Diabetes, Shanghai Diabetes Institute, Shanghai Key Laboratory of Diabetes Mellitus, Shanghai, China.

Aims: Currently, there is a lack of evidence regarding time in tight range (TITR) and long-term adverse outcomes. We aimed to investigate the association between TITR and the risk of all-cause and cardiovascular mortality among patients with type 2 diabetes.

Materials And Methods: A total of 6061 patients with type 2 diabetes were prospectively recruited in a single centre.

View Article and Find Full Text PDF

Aims: The aim of this study was to assess postprandial glycaemic outcomes using automated insulin delivery with faster acting insulin aspart (FIA) or standard insulin aspart (SIA) over 4 weeks in youth (aged 10-18 years) with type 1 diabetes.

Materials And Methods: We undertook a secondary analysis of postprandial glycaemic outcomes from a double-blind, randomised, crossover study comparing FIA to SIA using an investigational version of MiniMed™ 780G. Endpoints included postprandial time in tight range (70-140 mg/dL; TITR), postprandial glucose excursions and peak glucose, and incremental area under curve (iAUC).

View Article and Find Full Text PDF

Dual physiological responsive structural color hydrogel particles for wound repair.

Bioact Mater

April 2025

Joint Centre of Translational Medicine, Wenzhou Key Laboratory of Interdiscipline and Translational Medicine, The First Affiliated Hospital of Wenzhou Medical University, Wenzhou, China.

Hydrogel-based patches have demonstrated their values in diabetic wounds repair, particularly those intelligent dressings with continuous repair promoting and monitoring capabilities. Here, we propose a type of dual physiological responsive structural color particles for wound repair. The particles are composed of a hyaluronic acid methacryloyl (HAMA)-sodium alginate (Alg) inverse opal scaffold, filled with oxidized dextran (ODex)/quaternized chitosan (QCS) hydrogel.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!