Accuracy assessment of catheter guidance technology in electrophysiology procedures: a comparison of a new 3D-based fluoroscopy navigation system to current electroanatomic mapping systems.

J Cardiovasc Electrophysiol

The Lucas Center, Department of Radiology, Stanford School of Medicine, Stanford, California, USA; Department of Cardiovascular Diseases (partner site Munich Heart Alliance), German Heart Centre, Munich, Germany.

Published: January 2014

Background: With increasing complexity in electrophysiology (EP) procedures, the use of electroanatomic mapping systems (EAMS) as a supplement to fluoroscopy has become common practice. This is the first study that evaluates spatial and point localization accuracy for 2 current EAMS, CARTO3(®) (Biosense Webster, Diamond Bar, CA, USA) and EnSite Velocity(®) (St. Jude Medical Inc., St. Paul, MN, USA), and for a novel overlay guidance (OG) software (Siemens AG, Forchheim, Germany) in a phantom experiment.

Methods And Results: A C-arm CT scan was performed on an acrylic phantom containing holes and location markers. Spatial accuracy was assessed for each system using distance measurements involving known markers inside the phantom and properly placed catheters. Anatomical maps of the phantom were acquired by each EAMS, whereas the 3D-based OG software superimposed an overlay image of the phantom, segmented from the C-arm CT data set, onto biplane fluoroscopy. Registration processes and landmark measurements quantitatively assessed the spatial accuracy of each technology with respect to the ground truth phantom. Point localization performance was 0.49 ± 0.25 mm in OG, 0.46 ± 0.17 mm in CARTO3(®) and 0.79 ± 0.83 mm in EnSite(®) . The registration offset between virtual visualization and reality was 1.10 ± 0.52 mm in OG, 1.62 ± 0.77 mm in CARTO3(®) and 2.02 ± 1.21 mm in EnSite(®) . The offset to phantom C-arm CT landmark measurements was 0.30 ± 0.26 mm in OG, 0.24 ± 0.21 mm in CARTO3(®) and 1.32 ± 0.98 mm in EnSite(®) .

Conclusions: Each of the evaluated EP guidance systems showed a high level of accuracy; the observed offsets between the virtual 3D visualization and the real phantom were below a clinically relevant threshold of 3 mm.

Download full-text PDF

Source
http://dx.doi.org/10.1111/jce.12264DOI Listing

Publication Analysis

Top Keywords

electrophysiology procedures
8
electroanatomic mapping
8
mapping systems
8
point localization
8
phantom
8
spatial accuracy
8
landmark measurements
8
virtual visualization
8
accuracy
5
accuracy assessment
4

Similar Publications

Objectives: The objective of this study was to assess the quality of ECG recordings and the concordance between the automatic detection of cardiac arrhythmia episodes by a patch ECG and an insertable cardiac monitor.

Design: Prospective cohort study.

Setting And Participants: Endurance athletes diagnosed with paroxysmal atrial fibrillation (AF) and no other relevant comorbidities participating in a randomised controlled trial on the effects of training adaption.

View Article and Find Full Text PDF

Adipose ZFP36 protects against diet-induced obesity and insulin resistance.

Metabolism

January 2025

State Key Laboratory for Innovation and Transformation of Luobing Theory, Key Laboratory of Cardiovascular Remodeling and Function Research, Chinese Ministry of Education, Chinese National Health Commission and Chinese Academy of Medical Sciences, Department of Cardiology, Qilu Hospital of Shandong University, Jinan, China. Electronic address:

Aims: Obesity, as a worldwide healthcare problem, has become more prevalent. ZFP36 is a well-known RNA-binding protein and involved in the posttranscriptional regulation of many physiological processes. Whether the adipose ZFP36 plays a role in obesity and insulin resistance remains unclear.

View Article and Find Full Text PDF

Objective: Cognitive deficits are one of the most debilitating comorbidities in epilepsy and other neurodegenerative, neuropsychiatric, and neurodevelopmental brain disorders. Current diagnostic and therapeutic options are limited and lack objective measures of the underlying neural activities. In this study, electrophysiological biomarkers that reflect cognitive functions in clinically validated batteries were determined to aid diagnosis and treatment in specific brain regions.

View Article and Find Full Text PDF

Background: Stroke associated with atrial fibrillation (AF) is a significant cause of mortality. This study analyzed demographic trends and disparities in mortality rates due to stroke in AF patients aged ≥25 years.

Methods: A retrospective analysis was conducted to acquire death data using the Centers for Disease Control and Prevention database from 1999 to 2020.

View Article and Find Full Text PDF

Introduction: Ultrasound is important in heart diagnostics, yet implementing effective cardiac ultrasound requires training. While current strategies incorporate digital learning and ultrasound simulators, the effectiveness of these simulators for learning remains uncertain. This study evaluates the effectiveness of simulator-based versus human-based training in Focused Assessed with Transthoracic Echocardiography (FATE).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!