Background: Spinal cord stimulation (SCS) is an accepted treatment in patients with failed back surgery (FBS), complex regional pain syndrome (CRPS) and persistent radicular pain following surgery. In order to avoid patient hazards or device malfunction manufacturers advise to abstain from magnetic resonance imaging (MRI) in patients with implanted electrodes or pulse generators.

Methods: In a prospective study, 13 patients harbouring an implanted Medtronic Spinal Cord Stimulation (SCS) device underwent MRI (1.5 T) of the lumbar (n=13), the cervical (n=2) or the thoracic spine (n=1) following the development of new spinal symptoms. An adapted MRI protocol was used limiting the transmitted energy and specific absorption rate. Tolerability and safety were assessed by means of a standardized patient evaluation form documenting pain on a visual analogue scale (0-10), neurologic deficit, and discomfort during the scan. In addition, overall satisfaction with the examination procedure was rated on a Likert scale (1-5). Image quality was rated independently and blinded to the presence of a SCS device by the radiologist and the surgeon as equivalent, superior or inferior compared to the standard spine MRI examination.

Results: None of the 13 patients investigated by the modified spinal MRI protocol experienced new neurological deficits, worsening of symptoms or a defect/malfunction of the implant device. Three patients (23.1 %) reported transient warm sensation in the location of the electrode and in one case intermittent slight tingling in the lower extremities. Overall satisfaction with the examination was 1.13 ± 0.34 according to Likert scale (1-5). The image quality was rated - not statistically significant - slightly inferior to standard lumbar spine imaging (0.82 ± 0.54) with a kappa value of 0.68 between the two investigators. MRI examinations detected relevant and new lesions in 9 (69.2 %) patients which affected treatment in 8 (61.5 %) individuals.

Conclusion: Using a protocol with a reduced specific energy absorption rate, spinal MRI examinations in patients with SCS can be considered safe. The current view that neurostimulators are a general contraindication to MR examinations has to be reconsidered in patients with new or progressive spinal symptoms.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00701-013-1885-8DOI Listing

Publication Analysis

Top Keywords

absorption rate
12
spinal cord
12
cord stimulation
12
image quality
12
patients
9
spinal
8
magnetic resonance
8
resonance imaging
8
reduced specific
8
specific absorption
8

Similar Publications

Ecological concrete by partially substitution of cement with Cameroonian corn stover ash.

Heliyon

January 2025

Mechanics Laboratory, Doctoral Training Unit in Engineering Sciences, Doctoral School of Fundamental and Applied Sciences, University of Douala, P.O. Box: 2701, Douala, Cameroon.

This study focuses on the influence of the partial substitution of cement by Cameroonian corn stover ash (CCSA) on the physical and mechanical behavior of concrete. For this, as materials used, one has first the corn stovers coming from the Bandjoun town in the Koung-khi division, in the West region of Cameroon, which are used to obtain the ashes, while the sand used, came from the Sanaga River in the coastal region of Cameroon. In order to obtain the CCSA, the corn stover is calcined in an oven at 600 °C for 6 h and then characterized; the characterization included infrared spectrometry, X-ray fluorescence spectrometry, fineness of grinding, and absolute density.

View Article and Find Full Text PDF

Background: The process of aging is associated with the deterioration of normal body functions as well as lower health awareness due to a reduced ability to acquire and understand health-related information. The subjects of this study were older adults with diabetes mellitus living in community settings. Visual-learning-based Steno diabetes dialogue cards were used to strengthen the learning and knowledge absorption capabilities of the subjects to improve diabetes health knowledge and increase compliance with blood sugar control protocols to achieve blood sugar control goals and delay / reduce the severity of complications.

View Article and Find Full Text PDF

Using light to image millimeter wave based on stacked meta-MEMS chip.

Light Sci Appl

January 2025

Institute of Modern Optics, Nankai University, Tianjin Key Laboratory of Micro-scale Optical Information Science and Technology, Tianjin, 300350, China.

A stacked metamaterial MEMS (meta-MEMS) chip is proposed, which can perfectly absorb electromagnetic waves, convert them into mechanical energy, drive movement of the optical micro-reflectors array, and detect millimeter waves. It is equivalent to using visible light to image a millimeter wave. The meta-MEMS adopts the design of upper and lower chip separation and then stacking to achieve the "dielectric-resonant-air-ground" structure, reduce the thickness of the metamaterial and MEMS structures, and improve the performance of millimeter wave imaging.

View Article and Find Full Text PDF

Impact of aerosol-photolysis interaction on the ozone concentration in the upper boundary layer on Mountain Everest.

Sci Total Environ

January 2025

SKL-ESPC & SEPKL-AERM, College of Environmental Sciences and Engineering and Center for Environment and Science, Peking University, Beijing 100871, China.

Previous research has revealed that, during the late afternoon, the ozone (O3) concentration tends to elevate at the northern perimeter of Mount Everest (5200 m above sea level). This increase is attributed to the natural gradient of rising O3 concentration with height, exacerbated by the corresponding downstream mountain winds. Our recent field observations corroborate this finding, showing a consistent increase in O3 concentrations by approximately 13.

View Article and Find Full Text PDF

Long-range exciton transport is crucial for optoelectronic devices based on organic semiconductors, but the method for increasing and regulating the exciton transport rate in organic semiconductors is still underexplored. Here we have achieved rapid-transporting excitons in organic crystals assembled of difluoroboron β-diketonate (BCZ) and found that the exciton transport rate of BCZ crystals can be regulated by the molecular packing form. Using transient absorption microscopy, we find that the BCZ-Y crystal in which BCZ molecules experience uniform head-to-tail antiparallel molecular packing has anisotropic long-range exciton transport.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!