Objective: We performed a multicenter study to calculate the upper reference limits (URL) for urine particle quantification in mid-stream samples by using automated urine analyzers.

Design & Methods: Two laboratories tested 283 subjects using a Sysmex UF-100, two other laboratories tested 313 subjects using Sysmex UF-1000i, whereas two other laboratories tested 267 subjects using Iris IQ®200.

Results: The URLs of UF-100 in females and males were 7.8/μL and 6.7/μL for epithelial cells (EC), 11.1/μL and 9.9/μL for red blood cells (RBC), 10.2/μL and 9.7/μL for white blood cells (WBC), and 0.85/μL and 0.87/μL for cylinders (CAST). The URLs of UF-1000i in females and males were 7.6/μL and 7.1/μL for EC, 12.2/μL and 11.1/μL for RBC, 11.9/μL and 11.7/μL for WBC, and 0.88/μL and 0.86/μL for CAST. The URLs of Iris IQ®200 in females and males were 7.8/μL and 6.6/μL for EC, 12.4/μL and 10.1/μL for RBC, 10.9/μL and 9.9/μL for WBC, and 1.1/μL and 1.0/μL for CAST.

Conclusion: The URLs obtained in this study were comparable to the lowest values previously reported in the literature. Moreover, no gender-related difference was observed, and analyzer-specific upper reference limits were very similar.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.cca.2013.09.012DOI Listing

Publication Analysis

Top Keywords

laboratories tested
12
females males
12
multicenter study
8
upper reference
8
reference limits
8
subjects sysmex
8
males 78/μl
8
blood cells
8
cast urls
8
gender's equality
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!