Objective: To explore the image quality of different tube voltage on digital chest radiograph for contrast detail phantom (CDRAD2.0) and occupational exposed workers.
Methods: The CDRAD2.0 phantom DR images of high KV and different tube voltages were analyzed by 3 readers, image quality figure (IQF) were calculated and compared; at the same time 136 exposed workers were examined with high-kV and DR chest radiograph of different tube voltages. Contrast to high-KV images, 10 anatomic sites were scored .The image differences were compared between DR and high-kV.
Results: On CDRAD2.0 phantom, the IQF value of DR images in 3 readers were minimum in the condition of 120 kV, average value was 22.25. The analysis of variance in model with random effects, the mean IQF value of different tube voltage DR image had a significant difference (F = 13.775, P<0.01); By Dunnett t-tests analysis, the mean IQF value of DR image in 120 kV and high kilovoltage had no difference (t = -0.58, P = 0.979); On clinical cases, the DR image of 120 kV showed the closest anatomy to the high KV, the mean had no significant difference with 0 (P > 0.05) with single sample t test.
Conclusion: On the CDRAD2.0 phantom or clinical exposed workers, the DR image quality of 120 kV tube voltage equals to high-KV basically.
Download full-text PDF |
Source |
---|
Appl Radiat Isot
January 2025
School of Artificial Intelligence, Wenzhou Polytechnic, Wenzhou, 325035, China. Electronic address:
For the purpose of assessing image quality and calculating patient X-ray dosage in radiology, computed tomography (CT), fluoroscopy, mammography, and other fields, it is necessary to have prior knowledge of the X-ray energy spectrum. The main components of an X-ray tube are an electron filament, also known as the cathode, and an anode, which is often made of tungsten or rubidium and angled at a certain angle. At the point where the electrons generated by the cathode and the anode make contact, a spectrum of X-rays with energies spanning from zero to the maximum energy value of the released electrons is created.
View Article and Find Full Text PDFDiagnostics (Basel)
December 2024
Postgraduation School in Radiodiagnostics, Università degli Studi di Milano, Via Festa del Perdono 7, 20122 Milan, Italy.
Dynamic digital radiography (DDR) is a recent imaging technique that allows for real-time visualization of thoracic and pulmonary movement in synchronization with the breathing cycle, providing useful clinical information. A 46-year-old male, a former smoker, was evaluated for unexplained dyspnea and reduced exercise tolerance. His medical history included a SARS-CoV-2 infection in 2021.
View Article and Find Full Text PDFRadiat Environ Biophys
January 2025
Department of Physics, Yazd University, Yazd, Iran.
Polymer nanocomposites have been investigated as lightweight and suitable alternatives to lead-based clothing. The present study aims to fabricate flexible, lead-free, X-ray-shielding composites using a polyvinyl chloride (PVC) matrix and different nanostructures. Four different nanostructures containing impure tungsten oxide, tungsten oxide (WO), barium tungstate (BaWO), and bismuth tungstate (BiWO) were synthesized through various methods.
View Article and Find Full Text PDFPurpose: With the widespread introduction of dual energy computed tomography (DECT), applications utilizing the spectral information to perform material decomposition became available. Among these, a popular application is to decompose contrast-enhanced CT images into virtual non-contrast (VNC) or virtual non-iodine images and into iodine maps. In 2021, photon-counting CT (PCCT) was introduced, which is another spectral CT modality.
View Article and Find Full Text PDFDiagnostics (Basel)
December 2024
Medical Diagnostic Imaging Department, College of Health Science, University of Sharjah, Sharjah 27272, United Arab Emirates.
: Computed Tomography (CT) is crucial in medical diagnosis, particularly for head examinations. Diagnostic Reference Levels (DRLs) are pivotal in balancing diagnostic efficacy with radiation safety. International organizations such as the International Atomic Energy Agency (IAEA) and the International Commission on Radiological Protection (ICRP) provide guidelines for establishing DRLs, emphasizing their importance in optimizing radiation doses.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!