Purpose: To develop with a thin film interferometer new parameters to describe the drying properties of contact lenses; these are: time to first break-up (onset latency), duration of lens surface drying (drying duration), maximum speed of increase in the drying area (maximum speed), and the time to reach maximum drying speed (peak latency). These new parameters were compared with the contact angle (CA) measurement of contact lenses by the captive bubble (CB) technique.
Methods: A thin film interferometer was connected to a digital camera, which captured images of the pre-lens liquid film of seven soft contact lenses in vitro: 1-Day Acuvue Moist, ACUVUE Oasys (Johnson & Johnson Vision Care); SofLens 38, PureVision (Bausch & Lomb); AirOptix Night & Day, AirOptix Aqua (CIBA Vision, Corp); and Proclear (Cooper Vision). The images were obtained from the lens surface when it was wet until it became dry (7 images per second) to generate video clips. A program was created in MATLAB to analyze the results. The CAs of the lens surfaces were measured by the CB technique with an OCA-20 contact angle analyzer (Data Physics Instruments).
Results: There were no significant relationships found between the CA and the new drying parameters (P > 0.05). Analysis of the results using analysis of variance and post hoc tests showed that the surface drying dynamics measures effectively differentiated between more of the soft lenses than the CA.
Conclusions: A new technique is described for assessing the drying characteristics of contact lenses by observations of the pattern, size, and speed of pre-lens liquid film break-up observed by a thin film interferometry. The interferometry method allowed evaluation and differentiation between lens materials. This new technique has the major advantage that it can be applied to evaluate contact lens wettability in vivo and in vitro.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/ICL.0b013e318297394c | DOI Listing |
Polymers (Basel)
January 2025
Department of Optometry & Vision Science, Daegu Catholic University, Gyeongsan 38430, Republic of Korea.
This study aims to build an optimal drug delivery system by manufacturing and evaluating a hydrogel contact lens using Tretinoin (ATRA) and protein nanoparticles to improve the drug delivery system as an ophthalmic medical contact lens. To evaluate the optical and physical properties of the manufactured lens, the spectral transmittance, refractive index, water content, contact angle, AFM, tensile strength, drug delivery, and antibacterial properties were analyzed. The contact lens was manufactured to contain ATRA and bovine serum albumin (BSA) in different ways, and the results confirmed that A, B, and C each had different physical properties.
View Article and Find Full Text PDFHealthcare (Basel)
January 2025
Eye Institute and Department of Ophthalmology, Eye & ENT Hospital, Fudan University, Shanghai 200031, China.
This study aimed to analyze the research progress on the use of a multifocal lens for myopic control throughout the 21st century, utilizing bibliometric analysis. Publications related to multifocal lenses from 2001 to 2024 were searched on the Web of Science core collection (WoSCC) database. VOSviewer (Version 1.
View Article and Find Full Text PDFOphthalmic Physiol Opt
January 2025
ISEC LISBOA-Instituto Superior de Educação e Ciências, Lisbon, Portugal.
Purpose: The purpose of this study was to compare the perception and understanding of the information provided by ChatGPT regarding myopia among optometry students, optometrists undertaking a Master degree and practicing optometrists.
Methods: This was a cross-sectional descriptive study using a structured questionnaire distributed via Wooclap to 225 participants (125 optometry students, 21 Masters students and 79 practicing optometrists). All participants evaluated the responses generated by ChatGPT Version 4.
Biosensors (Basel)
January 2025
Department of Optometry, Seoul National University of Science and Technology, 232 Gongneung-ro, Nowon-gu, Seoul 01811, Republic of Korea.
Ocular cystinosis is a disease in which accumulated cystine crystals cause damage to the eyes, necessitating timely treatment and ongoing monitoring of cystine levels. The current treatment involves frequent administration of cysteamine eye drops, which suffer from low bioavailability and can lead to drug toxicity, making it essential to prescribe an appropriate dosage based on the patient's condition. Additionally, cystine crystal levels are typically assessed subjectively via slit-lamp examination, requiring frequent clinical visits and causing discomfort for the patient.
View Article and Find Full Text PDFJ Contact Lens Res Sci
July 2024
Illinois College of Optometry, Chicago.
Background And Objective: This study determined whether practitioners specializing in keratoconus (KC) adhere to published guidelines for disease management and to what extent comorbid conditions of dry eye, contact lens tolerance, and psychological consequences of KC are formally assessed as part of long-term management.
Materials And Methods: This cross-sectional study used an IRB-approved, Internet-based, REDCap platform. Descriptive statistics are presented.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!