Background: Implantable cardioverter defibrillator (ICD) therapy has increased in Western countries. The aim of the study was to investigate the function of ICDs using a human-shaped phantom in electric and magnetic fields of 400 kV power lines.

Methods: The phantom was used in the following manner: isolated from the ground, earthed from a foot, or earthed from a hand.

Results: We performed 37 ICD tests using 10 different ICD devices. When the electric fields varied from 6.8 kV/m to 7.5 kV/m (humidity 70.5%) and the magnetic field was 2.0 μT, one of the ICDs tested recorded 258 ventricular beats/min when a simulated heart signal was applied to ICD electrodes. When the exposure was 5.1 kV/m, the same ICD had a similar disturbance; however, in a 0.9 kV/m field, it worked correctly.

Conclusions: Consequently, no effect on ICDs functioning was observed up to 0.9 kV/m, while anomalous behavior in some conditions was observed when levels exceeded 5.1 kV/m; ICD malfunctioning seems possible within 11.5 m from 400 kV power lines or in conditions inducing exposures exceeding 5 kV/m. Further development of this research field is needed.

Download full-text PDF

Source
http://dx.doi.org/10.1111/pace.12270DOI Listing

Publication Analysis

Top Keywords

400 power
12
implantable cardioverter
8
electric magnetic
8
magnetic fields
8
fields 400
8
power lines
8
kv/m icd
8
kv/m
7
icd
6
cardioverter defibrillators
4

Similar Publications

Technology-critical elements (TCEs) refer to the elements that play an important role in many emerging technologies and the production of advanced materials, and these include lanthanides, tungsten and vanadium. Actinides, Tl, and Pb, which also belong to TCEs, are abundantly used in power generation, industrial applications, and modern agricultural practices. The information on the influence of these elements on the aquatic environment and biota is still rather scarce.

View Article and Find Full Text PDF

Microwave-assisted extraction of pectin from Dillenia indica (DI) fruit was optimized using Box-Behnken design to maximize yield and quality. Parameters such as solid:solvent (1:10-1:30), microwave power (200-600 W), and extraction time (4-10 min) were varied to determine the optimal conditions. Through experimentation, the optimized extraction parameters were identified as 1:23.

View Article and Find Full Text PDF

Investigation of the Ocular Response and Corneal Damage Threshold of Exposure to 28 GHz Quasi-millimeter Wave Exposure.

Health Phys

January 2025

Division of Vision Research for Environmental Health, Medical Research Institute and Department of Ophthalmology, Kanazawa Medical University, Kahoku, Japan.

Electromagnetic radiation energy at millimeter wave frequencies, typically 30 GHz to 300 GHz, is ubiquitously used in society in devices for telecommunications; radar and imaging systems for vehicle collision avoidance, security screening, and medical equipment; scientific research tools for spectroscopy; industrial applications for non-destructive testing and precise measurement; and military and defense applications. Understanding the biological effects of this technology is essential. We have been investigating ocular responses and damage thresholds comparing various frequencies using rabbit eyes and dedicated experimental apparatus.

View Article and Find Full Text PDF

Perfluorooctanesulfonic acid (PFOS) is one of the most investigated Per- and polyfluoroalkyl substances (PFAS) for being the strongest compound to eliminate and having adverse health concerns. In this work, we have conducted the sonochemical treatment of PFOS simulated water under high (500 kHz) and low (22 kHz) frequencies while monitoring the operational parameters via an integrated sonochemical system. The integrated advanced sonochemical system includes software to monitor treatment power, solution temperature and frequency while allowing distinctive control of the reaction conditions.

View Article and Find Full Text PDF

A 47-year-old woman with a 12-year history of anemia and high C-reactive protein (CRP) levels was admitted to our hospital with worsening fatigue and night sweats. She had high levels of immunoglobulin G (IgG; 4182 mg/dL), IgA (630.6 mg/dL), and CRP (7.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!