Enhancement cone beam computed tomography filters improve in vitro periimplant dehiscence detection.

Oral Surg Oral Med Oral Pathol Oral Radiol

Department of Clinical Dentistry, School of Dentistry, Federal University of Espírito Santo, Vitória, Espírito Santo, Brazil. Electronic address:

Published: November 2013

Objective: To investigate whether cone beam computed tomography filters would improve periimplant dehiscence detection.

Study Design: A hundred titanium implants were placed in bovine ribs in which defects simulating periimplant dehiscence had previously been created. After images acquisition, three oral radiologists assessed them with and without the following filters: Angio Sharpen high 5 × 5, Shadow, Sharpen 3 × 3, Sharpen Mild, and Smooth. The McNemar test verified the disagreement between all images versus the reference standard and original images versus images with filters; P < .05 was considered statistically significant.

Results: Dehiscence detection using the original images and the Shadow filter disagreed from the reference standard (P < .05), as well as when using the filters instead of the original images (P < .05).

Conclusion: All the filters tested, with the exception of the Shadow, improved periimplant dehiscence detection. The Sharpen 3 × 3 filter was considered best for this task.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.oooo.2013.06.029DOI Listing

Publication Analysis

Top Keywords

periimplant dehiscence
16
dehiscence detection
12
original images
12
cone beam
8
beam computed
8
computed tomography
8
tomography filters
8
filters improve
8
images versus
8
reference standard
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!