Heat fixed but unstained slide smears are infectious to laboratory staff.

Indian J Tuberc

Division of Clinical Microbiology, Department of Laboratory Medicine, All India Institute of Medical Sciences, New Delhi, India.

Published: July 2013

Background And Aim: In a clinical microbiology laboratory, heat fixed slide smears are commonly transported from one place to another for staining with different stains and also for onsite proficiency testing of laboratory technicians for accreditation of the laboratories. These smears are frequently handled without gloves by the staff in developing countries. Therefore, this study was conducted to check the survivability of tubercle bacilli on smears after physical and chemical treatments.

Methods: A total of 196 AFB positive smears were analyzed. Of these, 116 were stained with Ziehl Neelsen (ZN), 60 with cold Kinyoun and 10 were unstained but heat fixed and 10 were neither stained nor heat fixed. The last 20 smears served as controls. The ZN and Kinyoun stained smears were 0-1.5-year-old and stored at room temperature in slide boxes, while control smears were freshly prepared. All smears were prepared from sputum samples positive for acid fast bacilli. All four sets were subjected to slide culture to see if mycobacteria could survive and grow in any. For slide culture, a new and safe device was used, which is designed for three in one purpose: cell cultivation, direct observation of the growth under microscope and cell harvesting inside the closed tube. The slide smears were directly dipped into this tube that contained liquid culture medium. The tubes were incubated at 37 degreeC for four weeks. The growth, if any, was confirmed by MPT-64 rapid test and subculture on LJ slants.

Results: No growth was observed in ZN and Kinyoun stained slide smears. However, significant growth was observed in both control sets; the unstained non heat fixed as well as heat fixed slide smears.

Conclusions: The results of our study indicate that tubercle bacilli remain viable even after heat fixation and carry risk of infection by contact However, stained smears are safe for handling and storage.

Download full-text PDF

Source

Publication Analysis

Top Keywords

heat fixed
24
slide smears
16
smears
12
slide
8
fixed slide
8
tubercle bacilli
8
unstained heat
8
kinyoun stained
8
stained smears
8
slide culture
8

Similar Publications

Purpose: Exercise-induced heat acclimation can mitigate age-related reductions in heat-loss capacity, though performing repeated bouts of strenuous exercise in the heat may be untenable for many older adults. While short-term passive heat acclimation (e.g.

View Article and Find Full Text PDF

Coral reefs worldwide are threatened by increasing ocean temperatures because of the sensitivity of the coral-algal symbiosis to thermal stress. Reef-building corals form symbiotic relationships with dinoflagellates (family Symbiodiniaceae), including those species which acquire their initial symbiont complement predominately from their parents. Changes in the composition of symbiont communities, through the mechanisms of symbiont shuffling or switching, can modulate the host's thermal limits.

View Article and Find Full Text PDF

Advanced adiabatic compressed air energy storage systems dynamic modelling: Impact of the heat storage device.

Heliyon

January 2025

IFP Energies nouvelles, 1 et 4 avenue de Bois Préau, 92852, Rueil-Malmaison, France.

Advanced Adiabatic Compressed Air Energy Storage (AACAES) is a technology for storing energy in thermomechanical form. This technology involves several equipment such as compressors, turbines, heat storage capacities, air coolers, caverns, etc. During charging or discharging, the heat storage and especially the cavern will induce transient behavior of operating points, notably temperature, pressure, and volume flow.

View Article and Find Full Text PDF

: In this paper, we investigate the association of glycoprotein 96 (GP96) and androgen receptor (AR) expression with clinicopathological factors, additional axillary lymph node burden, and their potential role in predicting 5-year overall survival (OS) and disease-free survival (DFS) in breast cancer (BC) patients with sentinel lymph node (SLN) involvement. We also explore the prognostic value of the presence of extranodal extension (ENE) in SLN. : We retrospectively enrolled 107 female patients with cT1-T2 invasive BC and positive SLN biopsy.

View Article and Find Full Text PDF

Effect of uniaxial bending methods on the flexural strength and Weibull analysis of heat-polymerized, CAD/CAM milled, and 3D-printed denture base resins.

Dent Mater

January 2025

Department of Oral Technology, Medical Faculty, University Hospital Bonn, Bonn, North Rhine-Westphalia, Germany; Department of Fixed Prosthodontics, Faculty of Dentistry, Suez Canal University, Ismailia, Egypt.

Objectives: To compare the flexural strength and modulus of denture base resins manufactured by conventional methods, 3-dimensional (3D) printing, and computer-aided design and computer-aided manufacturing (CAD/CAM) milling using 3-point bending (3PB) and 4-point bending (4PB) methods after simulated aging.

Methods: Ninety bars (64 ×10 ×3.3 mm) were prepared from heat-polymerized (Lucitone-199), CAD/CAM milled (G-CAM), and 3D-printed (Denturetec) denture base resins (n = 30 per material).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!