Conductance sizing balloon for measurement of peripheral artery minimal stent area.

J Vasc Surg

Department of Biomedical Engineering, Indiana University-Purdue University Indianapolis, Indianapolis, Ind; Department of Surgery, Indiana University-Purdue University Indianapolis, Indianapolis, Ind; Department of Cellular and Integrative Physiology, Indiana University-Purdue University Indianapolis, Indianapolis, Ind. Electronic address:

Published: September 2014

Background: Because stent underdeployment occurs frequently, accurate minimal stent area (MSA) measurement during postdilatation is necessary. This study investigated the accuracy and repeatability for MSA determination using a novel conductance balloon (CB) catheter for peripheral vessels.

Methods: The CB catheter is a standard balloon catheter that measures electrical conductance (ratio of current/voltage drop) in real-time during inflation, which directly relates to the balloon cross-sectional area through Ohm's law. CB measurements were made in 4- to 10-mm phantoms on the bench, ex vivo in stents fully deployed in diseased human peripheral arteries, and in vivo in stents fully deployed in peripheral vessels in six swine. CB measurement accuracy and repeatability were calculated and compared with the known dimension (bench phantoms) or with intravascular ultrasound (IVUS) measurement after stent deployment (ex vivo and in vivo).

Results: CB measurements were highly accurate (error: 1.8% bench, 5% ex vivo, and 5% in vivo) and repeatable (error: 0.9% bench, 1.8% ex vivo, and 1.3% in vivo), with virtually no bias (average difference in measurements: -0.05 mm bench CB vs known phantom diameters, -0.06 mm ex vivo CB vs IVUS, and -0.11 mm in vivo CB vs IVUS).

Conclusions: The CB sizing capability can be integrated within a standard balloon catheter (two-in-one function) to provide accurate, real-time assessment of MSA to ensure full stent apposition rather than the use of pressure as a surrogate for size.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jvs.2013.06.095DOI Listing

Publication Analysis

Top Keywords

balloon catheter
12
vivo
9
minimal stent
8
stent area
8
accuracy repeatability
8
standard balloon
8
bench vivo
8
vivo stents
8
stents fully
8
fully deployed
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!