Two-year follow-up of percutaneous coronary intervention using EucaTax or Cypher.

Cardiovasc Revasc Med

Russian Cardiology Research and Production Complex, Department of Atherosclerosis Problems. Electronic address:

Published: April 2014

Background: The EucaTax stent (EUPES) is a coronary stent with biodegradable polymer and camouflage coating that has been developed to promote the complete elution of drugs and decrease the risk of late complications. The aim of this study was to evaluate the efficacy and safety of the double-coated EUPES in patients with stable angina versus sirolimus-eluting stent CYPHER (SES) with permanent polymer coating.

Methods And Materials: The study included consecutive patient with at least 70% de novo coronary lesions in one or two native coronary arteries and who had undergone the coronary stenting using either EUPES or SES. We evaluated the 2-year major adverse cardiac events (MACE) rates, including total death (cardiac, non-cardiac), myocardial infarction (MI), target lesion revascularisation (TLR) and stent thrombosis.

Results: Between 2006 and 2009 this observational, prospective, single centre study included 602 patients (282 with EUPES and 320 with SES). At 2years, the rates of TLR (16.3% versus 6.25%; p=0.0001) and MACE (18.4% versus 7.8%; p=0.001) were significantly higher in the EUPES than in the SES group. The rate of TLR was significantly higher in the EUPES group compared with SES group in stenting of artery with a diameter less than 3mm, using stent length more than 18mm, as well as when the residual stenosis was more than 12%.

Conclusions: We found that EUPES was inferior to SES during the 2-year follow-up with respect to rates of MACE and TLR that were significantly higher in the EUPES than in the SES group.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.carrev.2013.07.005DOI Listing

Publication Analysis

Top Keywords

eupes ses
12
higher eupes
12
ses group
12
eupes
8
study included
8
tlr higher
8
ses
7
coronary
5
stent
5
two-year follow-up
4

Similar Publications

Objective: To compare the long-term results of transluminal balloon coronary angioplasty in patients after implantation of drug-eluting coronary artery stents with permanent or biosoluble polymer coatings.

Material And Methods: This was a prospective, comparative, single-center trial. It enrolled patients having at least 70% stenosis who had undergone intracoronary Cypher (SES) or EucaTAX (EUPES) stent implantation.

View Article and Find Full Text PDF

Two-year follow-up of percutaneous coronary intervention using EucaTax or Cypher.

Cardiovasc Revasc Med

April 2014

Russian Cardiology Research and Production Complex, Department of Atherosclerosis Problems. Electronic address:

Background: The EucaTax stent (EUPES) is a coronary stent with biodegradable polymer and camouflage coating that has been developed to promote the complete elution of drugs and decrease the risk of late complications. The aim of this study was to evaluate the efficacy and safety of the double-coated EUPES in patients with stable angina versus sirolimus-eluting stent CYPHER (SES) with permanent polymer coating.

Methods And Materials: The study included consecutive patient with at least 70% de novo coronary lesions in one or two native coronary arteries and who had undergone the coronary stenting using either EUPES or SES.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!