Background: Although stimulus evoked electromyography (EMG) is commonly used to confirm the accuracy of pedicle screw placement. There are no studies to differentiate between solid screws and hollow screws to the electrical resistance of pedicle screws. We speculate that the electrical resistance of the solid and hollow pedicle screws may be different and then a potential source of error with stimulus-evoked EMG may happen.
Materials And Methods: Resistance measurements were obtained from 12 pedicle screw varieties (6 screws of each manufacturer) across the screw shank based on known constant current and measured voltage. The voltage was measured 5 times at each site.
Results: Resistance of all solid screws ranged from 0.084 Ω to 0.151 Ω (mean =0.118 ± 0.024 Ω) and hollow screws ranged from 0.148 Ω to 0.402 Ω (mean = 0.285 ± 0.081 Ω). There was a significant difference of resistance between the solid screws and hollow screws (P < 0.05). The screw with the largest diameter no matter solid screws or hollow screws had lower resistance than screws with other diameters. No matter in solid screws group or hollow screws group, there were significant differences (P < 0.05) between the 5.0 mm screws and 6.0 mm screws, 6.0 mm screws and 7.0 mm screws, 5.0 mm screws and 7.0 mm screws, 4.5 mm screws and 5.5 mm screws, 5.5 mm screws and 6.5 mm screws, 4.5 mm screws and 6.5 mm screws. The resistance of hollow screws was much larger than the solid screws in the same diameter group (P < 0.05).
Conclusions: Hollow pedicle screws have the potential for high electrical resistance compared to the solid pedicle screws and therefore may affect the EMG response during stimulus-evoked EMG testing in pedicle screw fixation especially in minimally invasive percutaneous pedical screw fixation surgery.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3745688 | PMC |
http://dx.doi.org/10.4103/0019-5413.114915 | DOI Listing |
J Med Case Rep
January 2025
Bone and Joint Diseases Research Center, Shiraz University of Medical Sciences, Shiraz, Iran.
Background: Management of the extensive soft tissue injuries remains a significant challenge in orthopedic and plastic reconstructive surgery. Since the thumb is responsible for 40% of the functions of the hand, saving and reconstructing a mangled thumb is essential for the patient's future.
Case Presentation: This case report describes the management of a severe occupational thumb injury in a 25-year-old white Persian male who sustained an occupational injury to his left thumb, resulting in extensive burn, crush injury to the distal and proximal phalanx, and severe soft tissue damage to the first metacarpal, thenar, and palmar areas.
Chin J Traumatol
January 2025
Department of Emergency Surgery and Orthopaedic Surgery, Qilu Hospital of Shandong University, Jinan, 250012, China. Electronic address:
Purpose: The fixation method commonly employed worldwide for treating unstable fractures of the posterior pelvic ring is the percutaneous iliosacral screw technique. However, prolonged operation time and frequent fluoroscopies result in surgical risks. This study aimed to investigate whether a new triangulation method could reduce operative and fluoroscopy times and increase the accuracy of screw placement.
View Article and Find Full Text PDFJ Formos Med Assoc
January 2025
Department of Medical Imaging, National Taiwan University Hospital, Taipei, Taiwan; Department of Medical Imaging, National Taiwan University Cancer Center, Taipei, Taiwan.
Background And Purpose: Dual-energy computed tomography (DECT) pulmonary angiography can reliably detect cement pulmonary embolisms (CPEs) and parenchymal perfusion defects. This prospective observational study investigated CPEs in asymptomatic patients using DECT.
Methods: We enrolled 42 patients who underwent vertebroplasty or received cement screws for vertebral augmentation, examining them using spinal computed tomography and DECT pulmonary angiography.
Injury
January 2025
Department of Surgery, The Trauma and Orthopaedic Research Unit, The Canberra Hospital, Garran, Australian Capital Territory, Australia.
Background: Unstable posterior pelvic-ring fractures are rare and difficult to manage. There are many injury patterns, they are associated with high morbidity and mortality, and optimal surgical management remains contentions. This study aims to compare outcomes and complications for different surgical management of these injuries.
View Article and Find Full Text PDFInjury
January 2025
Department of Orthopaedic Surgery, Cedars - Sinai Medical Center, Los Angeles, CA, USA. Electronic address:
Objectives: The purpose of this study is to determine what demographic and anatomical variables affect successful placement of a superior medullary ramus screw, and how they affect the maximal diameter of that screw.
Methods: Design: Prognostic Level IV SETTING: Level I Trauma Center Patients/Participants: Two hundred consecutive patients underwent computed tomography (CT) of the pelvis. We included those patients aged 18 and older without osseous injury or abnormalities precluding measurement.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!