Aim: To assess the insertion procedure and performance of disc-retained gastrostomy tubes, recording complications and accidental displacements by prospective audit, and to determine whether primary placement of the tube off-licence was feasible.
Materials And Methods: Disc-retained 12 F single-lumen Monarch gastrostomy tubes (Enteral UK, Selby, UK) were inserted by three gastrointestinal interventional radiologists in a supra-regional cancer centre. The 12 F tubes required a 20 F peel-away sheath with four-point gastropexy fixation and were placed under conscious sedation, using electrocardiogram (EEG) bispectral index monitoring. Follow-up was performed in an in-house gastrostomy drop-in clinic at 1 week and 1 month, supplemented with weekly telephone follow-up. Patients also had open access to the gastrostomy drop-in clinic for immediate advice and complication management.
Results: Eighteen patients underwent primary insertion of a Monarch gastrostomy tube over 5 months. A total of 6/18 (33%) tubes displaced; 4/18 (22%) completely, 2/18 (11%) occult into the peritoneum. Four of 18 (22%) patients developed infection at the stoma site. Due to the unexpectedly poor performance of the tube, the study was terminated early.
Conclusion: Initial experience with the Monarch disc-retained gastrostomy tube demonstrates it unsuitable for primary placement with current protocols. In view of the potentially serious complications, the Medicines and Healthcare Products Regulatory Agency (MHRA) has been informed. A request has been made to the distributer to reassess the tube design and/or review the procedure promoted for primary placement.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.crad.2013.06.008 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!