Background: Desmosine/isodesmosine (DES/IDS) is a promising biomarker for estimating activity of elastin degradation. RESULTS/METHODOLOGY: A stable isotope dilution LC-MS/MS method for measuring serum/plasma DES/IDS was developed and validated. The reportable range of this assay was 0.1-160 ng/ml. Serum/plasma DES/IDS level was stable at room temperature or 4°C for 20 h, and for three freeze-thaw cycles. Interferences from endogenous compounds and ion suppression/enhancing effect were also evaluated. Our results suggest the absolute necessity of using an IS in the measurement. We found that serum/plasma DES/IDS levels from patients with chronic obstructive pulmonary disease and cystic fibrosis were significantly higher compared with healthy smokers.
Conclusion: These results demonstrate that the LC-MS/MS method provides sensitive, reproducible and accurate quantification of total serum/plasma DES/IDS.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.4155/bio.13.164 | DOI Listing |
Bioanalysis
August 2013
Translational Medicine Research Collaboration, TMRC Laboratory, James Arrott Drive, Dundee, DD1 9SY, UK.
Background: Desmosine/isodesmosine (DES/IDS) is a promising biomarker for estimating activity of elastin degradation. RESULTS/METHODOLOGY: A stable isotope dilution LC-MS/MS method for measuring serum/plasma DES/IDS was developed and validated. The reportable range of this assay was 0.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!