We demonstrate that spin-exchange dephasing of Larmor precession at near-Earth-scale fields is effectively eliminated by dressing the alkali-metal atom spins in a sequence of ac-coupled 2π pulses, repeated at the Larmor precession frequency. The contribution of spin-exchange collisions to the spectroscopic linewidth is reduced by a factor of the duty cycle of the pulses. We experimentally demonstrate resonant transverse pumping in magnetic fields as high as 0.1 G, present experimental measurements of the suppressed spin-exchange relaxation, and show enhanced magnetometer response relative to a light-narrowed scalar magnetometer.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6238214 | PMC |
http://dx.doi.org/10.1103/PhysRevLett.111.043002 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!