Background: The recently developed model-based iterative reconstruction (MBIR) enables significant reduction of image noise and artifacts, compared with adaptive statistical iterative reconstruction (ASIR) and filtered back projection (FBP). The purpose of this study was to evaluate lesion detectability of low-dose chest computed tomography (CT) with MBIR in comparison with ASIR and FBP.

Methods: Chest CT was acquired with 64-slice CT (Discovery CT750HD) with standard-dose (5.7 ± 2.3 mSv) and low-dose (1.6 ± 0.8 mSv) conditions in 55 patients (aged 72 ± 7 years) who were suspected of lung disease on chest radiograms. Low-dose CT images were reconstructed with MBIR, ASIR 50% and FBP, and standard-dose CT images were reconstructed with FBP, using a reconstructed slice thickness of 0.625 mm. Two observers evaluated the image quality of abnormal lung and mediastinal structures on a 5-point scale (Score 5 = excellent and score 1 = non-diagnostic). The objective image noise was also measured as the standard deviation of CT intensity in the descending aorta.

Results: The image quality score of enlarged mediastinal lymph nodes on low-dose MBIR CT (4.7 ± 0.5) was significantly improved in comparison with low-dose FBP and ASIR CT (3.0 ± 0.5, p = 0.004; 4.0 ± 0.5, p = 0.02, respectively), and was nearly identical to the score of standard-dose FBP image (4.8 ± 0.4, p = 0.66). Concerning decreased lung attenuation (bulla, emphysema, or cyst), the image quality score on low-dose MBIR CT (4.9 ± 0.2) was slightly better compared to low-dose FBP and ASIR CT (4.5 ± 0.6, p = 0.01; 4.6 ± 0.5, p = 0.01, respectively). There were no significant differences in image quality scores of visualization of consolidation or mass, ground-glass attenuation, or reticular opacity among low- and standard-dose CT series. Image noise with low-dose MBIR CT (11.6 ± 1.0 Hounsfield units (HU)) were significantly lower than with low-dose ASIR (21.1 ± 2.6 HU, p < 0.0005), low-dose FBP CT (30.9 ± 3.9 HU, p < 0.0005), and standard-dose FBP CT (16.6 ± 2.3 HU, p < 0.0005).

Conclusion: MBIR shows greater potential than ASIR for providing diagnostically acceptable low-dose CT without compromising image quality. With radiation dose reduction of >70%, MBIR can provide equivalent lesion detectability of standard-dose FBP CT.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3847631PMC
http://dx.doi.org/10.1186/1471-2342-13-27DOI Listing

Publication Analysis

Top Keywords

image quality
20
iterative reconstruction
16
image noise
12
low-dose mbir
12
low-dose fbp
12
standard-dose fbp
12
low-dose
11
image
9
fbp
9
adaptive statistical
8

Similar Publications

Olfactory Dysfunction in Allergic Rhinitis.

Clin Rev Allergy Immunol

December 2024

Department of Otolaryngology-Head and Neck Surgery, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China.

Olfactory dysfunction (OD) can have serious consequences as it hinders individuals from detecting important warning signals like smoke, spoiled food, and gas leaks. This can significantly impact their nutritional status, eating satisfaction, and overall quality of life. Allergic rhinitis (AR) is a common disease that greatly affects the quality of life and can lead to a decrease, distortion, or complete loss of olfactory ability.

View Article and Find Full Text PDF

Neuroendocrine tumors and diabetes mellitus: which treatment and which effect.

Endocrine

January 2025

Unit of Endocrinology, Department of Clinical and Molecular Medicine, ENETS Center of Excellence, Sapienza University of Rome, Rome, Italy.

Diabetes mellitus (DM) and neuroendocrine tumors (NET) can exert unfavorable effects on each other prognosis. In this narrative review, we evaluated the effects of NET therapies on glycemic control and DM management and the effects of anti-diabetic therapies on NET outcome and management. For this purpose, we searched the PubMed, Science Direct, and Google Scholar databases for studies reporting the effects of NET therapy on DM as well as the effect of DM therapy on NET.

View Article and Find Full Text PDF

Objectives: Automated breast ultrasound imaging (ABUS) results in a reduction in breast cancer stage at diagnosis beyond that seen with mammographic screening in women with increased breast density or who are at a high risk of breast cancer. It is unknown if the addition of ABUS to mammography or ABUS imaging alone, in this population, is a cost-effective screening strategy.

Methods: A discrete event simulation (Monte Carlo) model was developed to assess the costs of screening, diagnostic evaluation, biopsy, and breast cancer treatment.

View Article and Find Full Text PDF

Breast cancer is the most frequent non-dermatologic malignancy in women. Breast cancer is characterized by the expression of the human epidermal growth factor receptor type 2 (HER2), and the presence or lack of estrogen receptor (ER) and progesterone receptor (PR) expression. HER2 overexpression is reported in about 20 to 25% of breast cancer patients, which is usually linked to cancer progression, metastases, and poor survival.

View Article and Find Full Text PDF

Complexities of riverfront development for the hilly city of Paonta Sahib in India.

Environ Sci Pollut Res Int

January 2025

Water Resources Development and Management, Indian Institute of Technology, Roorkee, Uttarakhand, India.

The rapid urbanization, industrial growth, and socio-cultural activities along riverbanks in hilly cities are transforming land use and intensifying water infrastructure challenges. Paonta Sahib, a culturally significant town in Himachal Pradesh on the Yamuna River, along the foothills of the Himalayas exemplifies these pressures due to its religious tourism, industrialization, and mining activities. This study explores sustainable riverfront development at Paonta Sahib, addressing socio-cultural, environmental, and technical concerns essential for eco-sensitive urban planning.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!