Purpose: Attenuation correction for whole-body PET/MRI is challenging. Most commercial systems compute the attenuation map from MRI using a four-tissue segmentation approach. Bones, the most electron-dense tissue, are neglected because they are difficult to segment. In this work, the authors build on this segmentation approach by adding bones using a registration technique and assessing its performance on human PET images.

Methods: Twelve oncology patients were imaged with FDG PET/CT and MRI using a Turbo-FLASH pulse sequence. A database of 121 attenuation correction quality CT scans was also collected. Each patient MRI was compared to the CT database via weighted heuristic measures to find the "most similar" CT in terms of body geometry. The similar CT was aligned to the MRI with a deformable registration method. Two MRI-based attenuation maps were computed. One was a standard four-tissue segmentation (air, lung, fat, and lean tissue) using basic image processing techniques. The other was identical, except the bones from the aligned CT were added. The PET data were reconstructed with the patient's CT-based attenuation map (the silver standard) and both MRI-based attenuation maps. The relative errors of the MRI-based attenuation corrections were computed in 14 standardized volumes of interest, in lesions, and over whole tissues. The squared Pearson correlation coefficient was also calculated over whole tissues. Statistical testing was done with ANOVAs and paired t-tests.

Results: The MRI-based attenuation correction ignoring bone had relative errors ranging from -37% to -8% in volumes of interest containing bone. By including bone, the magnitude of the relative error was reduced in all cases (p<0.001), ranging from -3% to 4%. Further, the relative error in volumes of interest adjacent to bone was improved from a mean of -7.5% to 2% (p<0.001). In the other seven volumes of interest, including bone reduced the magnitude of relative error in three cases (p<0.001), had no effect in three cases, and increased relative error in one case. There was no statistically significant difference in the relative error in lesions. Over whole tissues, including bone slightly increased relative error in lung from 7.7% to 8.0% (p=0.002), in fat from 8.5% to 9.2% (p<0.001), and in lean tissue from -2.1% to 2.6% (p<0.001), but reduced the magnitude of relative error in bone from -14.6% to 1.3% (p<0.001). The correlation coefficient was essentially unchanged in all tissues regardless of whether bone was included or not.

Conclusions: The approach to include bones in MRI-based attenuation maps described in this work improves quantification of whole-body PET images in and around bony anatomy. The reduction in error is often large (tens of percents), and could alter image interpretation and subsequent patient care. Changes in other parts of the PET image are minimal and likely not of clinical significance.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.4816301DOI Listing

Publication Analysis

Top Keywords

attenuation correction
16
mri-based attenuation
16
attenuation
9
correction whole-body
8
whole-body pet/mri
8
attenuation map
8
four-tissue segmentation
8
segmentation approach
8
attenuation maps
8
relative errors
8

Similar Publications

Purpose: The common approach for organ segmentation in hybrid imaging relies on coregistered CT (CTAC) images. This method, however, presents several limitations in real clinical workflows where mismatch between PET and CT images are very common. Moreover, low-dose CTAC images have poor quality, thus challenging the segmentation task.

View Article and Find Full Text PDF

Purpose: Posterior fossa ring-enhancing lesions (PFREL) in the adult immunocompetent hosts pose a diagnostic challenge. We aimed to evaluate the spectrum of PFREL etiologies and propose a diagnostic algorithm.

Methods: This study involved a retrospective analysis of PFREL cases from our institution (January 2023 to April 2024) and a systematic literature review conducted using Embase and PubMed databases following the PRISMA 2020 guidelines.

View Article and Find Full Text PDF

Cadmium-cardiolipin disruption of respirasome assembly and redox balance through mitochondrial membrane rigidification.

J Lipid Res

January 2025

Physiology and Pathophysiology of Cells and Membranes, Medical School OWL, Bielefeld University, Bielefeld, Germany. Electronic address:

The environmental pollutant cadmium (Cd) poses a threat to human health through consumption of contaminated foodstuffs culminating in chronic nephrotoxicity. Mitochondrial dysfunction and excessive reactive oxygen species (ROS) are key to Cd cellular toxicity. Cd-lipid interactions have been less considered.

View Article and Find Full Text PDF

Coherent heterodyne lidars are typically used for windspeed and attenuated backscattering measurements. The lack of molecular backscattering detection capability has limited the calibrated backscattering measurements until recent advances in coherent lidar technology. In this work, the simultaneous detection of aerosol and molecular backscattering is demonstrated with coherent heterodyne lidar, and the results are compared with a state-of-the-art Raman lidar PollyXT as a reference in a long-range for the first time.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!