Ti2448 pedicle screw system augmentation for posterior lumbar interbody fusion.

Spine (Phila Pa 1976)

*Institute of Orthopedics, General Hospital of PLA, Beijing, China †Institute of Spine Surgery, Guangzhou Military Area General Hospital, Guangzhou, China; and ‡Institude of Orthopaedics, Xijing Hospital, the Fourth Military Medical University, Xi'an Shanxi Province, P. R. China.

Published: November 2013

AI Article Synopsis

  • A finite element analysis was conducted to assess the effectiveness of the Ti-24Nb-4Zr-7.9Sn (Ti2448) pedicle screw system in enhancing single-level posterior lumbar interbody fusion (PLIF).
  • The Ti2448 system showed decreased axial displacement at the fusion level compared to both screw-free status and the Ti-6Al-4V system, indicating potential benefits in maintaining spinal stability.
  • The study concluded that the Ti2448 screw system could reduce intradiscal pressures at adjacent discs, suggesting it may mitigate complications associated with traditional pedicle screw systems.

Article Abstract

Study Design: A finite element analysis was used.

Objective: To evaluate the feasibility of using the Ti-24Nb-4Zr-7.9Sn (Ti2448) pedicle screw system to augment single-level posterior lumbar interbody fusion (PLIF).

Summary Of Background Data: The Ti-6Al-4V pedicle screw system increases the risk of adjacent disc degeneration and stress-shielding effect due to enormous rigidity. A titanium alloy with much lower elastic modulus, Ti2448, may help to resolve the complications.

Methods: A finite element model of intact L3-S1 was established and then validated. Single-level PLIF at L4-L5 with or without a supplementary titanium-alloy pedicle screw system was simulated. A pure moment of 7.6 Nm and a 400 N preload was applied to the finite element model of PLIF, PLIF with the Ti-6Al-4V screw system, and PLIF with the Ti2448 screw system in flexion, extension, axial rotation, and lateral bending.

Results: The axial displacement at the fusion level decreased to 64%, 72%, 84%, and 92% of screw-free status in flexion, extension, axial rotation, and lateral bending, respectively, after augmentation of the Ti2448 screw system, which was 1% to 3% lower than the performance of the Ti-6Al-4V system. The angular displacement at the fusion level with the Ti2448 system was similar to that of the Ti-6Al-4V system, only 2% lower in flexion. Compared with the Ti-6Al-4V system, the Ti2448 system suppressed the increase of intradiscal pressures at the upper adjacent disc in all bending directions, but only in extension and axial rotation at the lower adjacent disc; the maximum stress experienced by cages and screws was higher in all bending directions when augmented with the Ti2448 system.

Conclusion: Using the Ti2448 screw system is suggested for augmenting single-level PLIF because it induces less disc intradiscal pressure at adjacent levels and the stress-shielding effect at implant-bone surface with stabilization performance compared with the Ti-6Al-4V screw system.

Download full-text PDF

Source
http://dx.doi.org/10.1097/BRS.0b013e3182a76fecDOI Listing

Publication Analysis

Top Keywords

screw system
36
pedicle screw
16
system
14
finite element
12
adjacent disc
12
ti2448 screw
12
extension axial
12
axial rotation
12
ti-6al-4v system
12
ti2448
9

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!