Background: The performance of a continuous glucose monitoring (CGM) system in the early stage of development was assessed in an inpatient setting that simulates daily life conditions of people with diabetes. Performance was evaluated at low glycemic, euglycemic, and high glycemic ranges as well as during phases with rapid glucose excursions.
Methods: Each of the 30 participants with type 1 diabetes (15 female, age 47 ± 12 years, hemoglobin A1c 7.7% ± 1.3%) wore two sensors of the prototype system in parallel for 7 days. Capillary blood samples were measured at least 16 times per day (at least 15 times per daytime and at least once per night). On two subsequent study days, glucose excursions were induced. For performance evaluation, the mean absolute relative difference (MARD) between CGM readings and paired capillary blood glucose readings and precision absolute relative difference (PARD), i.e., differences between paired CGM readings were calculated.
Results: Overall aggregated MARD was 9.2% and overall aggregated PARD was 7.5%. During induced glucose excursions, MARD was 10.9% and PARD was 7.8%. Lowest MARD (8.5%) and lowest PARD (6.4%) were observed in the high glycemic range (euglycemic range, MARD 9.1% and PARD 7.4%; low glycemic range, MARD 12.3% and PARD 12.4%).
Conclusions: The performance of this prototype CGM system was, particularly in the hypoglycemic range and during phases with rapid glucose fluctuations, better than performance data reported for other commercially available systems. In addition, performance of this prototype sensor was noticeably constant over the whole study period. This prototype system is not yet approved, and performance of this CGM system needs to be further assessed in clinical studies.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3879745 | PMC |
http://dx.doi.org/10.1177/193229681300700403 | DOI Listing |
Diabetes Metab Syndr Obes
January 2025
Department of Endocrinology, Shanghai Pudong Hospital, Fudan University, Shanghai, 201399, People's Republic of China.
Background: Direct Observation of Procedural Skills (DOPS) is a clinical assessment tool that enables trainers to observe medical students' procedural abilities in real-time clinical settings. It assesses students' knowledge application, decision-making, and skill proficiency during clinical tasks.
Methods: This study modifies the DOPS to evaluate the operation of insulin pumps (PUMP) and continuous glucose monitoring systems (CGMS) in diabetes management.
Langenbecks Arch Surg
January 2025
Department of Surgery, Division of HBP Surgery & Liver Transplantation, Korea University College of Medicine, Seoul, Korea.
Purpose: Pancreatectomy patients often experience challenging fluctuations in blood glucose levels; therefore, they require a reliable monitoring system. This study aimed to determine the accuracy and acceptability of a continuous glucose monitoring (CGM) system compared with the intermittent capillary glucose test in patients who have undergone pancreatectomy.
Methods: Thirty non-diabetic pancreatectomy patients participated.
Healthcare (Basel)
December 2024
Endocrinology Unit, Fondazione IRCCS Ca' Granda Ospedale Maggiore Policlinico, 20122 Milan, Italy.
The management of type 1 diabetes in pregnancy with new technologies is challenging. Sometimes the complexity of new-generation systems such as "continuous subcutaneous insulin infusion, CSII" and patient or provider preference do not allow their use, so women with type 1 diabetes in pregnancy continue to be treated with subcutaneous multiple-injection insulin therapy using pens. Smart insulin pens are new tools that allow for data collection on insulin dose and time of administration and have additional connectivity features.
View Article and Find Full Text PDFDiabetes Technol Ther
January 2025
Department of Metabolism, Digestion and Reproduction, Imperial College London, London, United Kingdom.
The accuracy of the latest generation Dexcom G7 sensors in individuals with diabetes undergoing hemodialysis has not previously been investigated. Participants with diabetes undergoing hemodialysis were recruited, with paired sensor glucose from Dexcom G7 recorded with plasma glucose analyzed in the laboratory, as well as the Freestyle Precision Pro glucometer and EKF Biosen C-Line analyzer. Ten adults (median age 64.
View Article and Find Full Text PDFSci Diabetes Self Manag Care
January 2025
Division of Endocrinology and Diabetes, Children's Hospital of Philadelphia, Philadelphia, Pennsylvania.
Purpose: The purpose of the study was to identify the most common reasons for and timing of continuous glucose monitoring (CGM) attrition in youth with type 1 diabetes (T1DM).
Methods: This single center retrospective chart review included youth with T1DM <22 years seen between November 1, 2021, and October 31, 2022. Data were gathered from CGM cloud-based software and the electronic medical record.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!