Objective: To develop and validate a simple, accurate HPTLC method for the analysis of 8-gingerol and to determine the quantity of 8-gingerol in Zingiber officinale extract and ginger-containing dietary supplements, teas and commercial creams.
Methods: The analysis was performed on 10×20 cm aluminium-backed plates coated with 0.2 mm layers of silica gel 60 F254 (E-Merck, Germany) with n-hexane: ethyl acetate 60: 40 (v/v) as mobile phase. Camag TLC Scanner III was used for the UV densitometric scanning at 569.
Results: This system was found to give a compact spot of 8-gingerol at retention factor (Rf) value of (0.39±0.04) and linearity was found in the ranges 50-500 ng/spot (r (2)=0.9987). Limit of detection (12.76 ng/spot), limit of quantification (26.32 ng/spot), accuracy (less than 2 %) and recovery (ranging from 98.22-99.20) were found satisfactory.
Conclusions: The HPTLC method developed for quantification of 8-gingerol was found to be simple, accurate, reproducible, sensitive and is applicable to the analysis of 8-gingerol in Zingiber officinale extract and ginger-containing dietary supplements, teas and commercial creams.