Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: Low-pressure hydrocephalus (LPH) is characterized by ventriculomegaly with persistent low intracranial pressure (ICP). Sub-zero drainage is needed for its management and multiple solutions have been described. Our aim is to report our experience with custom-made peritoneal catheters with larger inner diameter as an alternative treatment option.
Methods: We made a retrospective review of all patients diagnosed with LPH and treated with custom-made peritoneal catheters at the Virgen del Rocío Pediatric Neurosurgical Unit. Catheters were coated with antibiotic or silver. The inner diameter of ventricular catheters was 1.4 mm; peritoneal catheters were larger than usual (1.9 mm inner diameter).
Results: We identified four patients in whom five custom-made peritoneal catheters were used over a 3-year period. There were two males and the mean age was 10 years (6 months-17 years). In all patients, placement of an EVD was necessary for sub-zero drainage, with maximum negative pressure of -8 cm H20. The mean time of maintenance of EVD was 102 days (10 days-1 year). Finally, three ventriculoperitoneal (VP) valveless systems, one with antigravitation device, and one Pro-GAV VP shunt were placed, all of them with larger custom-made peritoneal catheters. After a mean follow-up period of 2.3 years (6 months-3 years), two patients are completely recovered, one patient is partially dependent for daily activities with good cognitive status, and the last one is a child who died due to his brain tumor.
Conclusion: The custom-made peritoneal catheters with larger inner diameter could be a good option for the management of this complex pathology.