Introduction: The purpose of this study was to analyze explanted PEEK rod spinal systems in the context of their clinical indications. We evaluated damage to the implant and histological changes in explanted periprosthetic tissues.

Methods: 12 patients implanted with 23 PEEK rods were revised between 2008 and 2012. PEEK rods were of the same design (CD Horizon Legacy, Medtronic, Memphis TN, USA). Retrieved components were assessed for surface damage mechanisms, including plastic deformation, scratching, burnishing, and fracture. Patient history and indications for PEEK rod implantation were obtained from analysis of the medical records.

Results: 11/12 PEEK rod systems were employed for fusion at one level, and motion preservation at the adjacent level. Surgical complications in the PEEK cohort included a small dural tear in one case that was immediately repaired. There were no cases of PEEK rod fracture or pedicle screw fracture. Retrieved PEEK rods exhibited scratching, as well as impressions from the set screws and pedicle screw saddles. PEEK debris was observed in two patient tissues, which were located adjacent to PEEK rods with evidence of scratching and burnishing.

Conclusion: This study documents the surface changes and tissue reactions for retrieved PEEK rod stabilization systems. Permanent indentations by the set screws and pedicle screws were the most prevalent observations on the surface of explanted PEEK rods.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3843804PMC
http://dx.doi.org/10.1007/s00586-013-2920-4DOI Listing

Publication Analysis

Top Keywords

peek rods
24
peek rod
20
peek
13
motion preservation
8
explanted peek
8
pedicle screw
8
retrieved peek
8
set screws
8
screws pedicle
8
rods
6

Similar Publications

Purpose: Implanted devices used in metastatic spine tumor surgery (MSTS) include pedicle screws, fixation plates, fixation rods, and interbody devices. A material to be used to fabricate any of these devices should possess an array of properties, which include biocompatibility, no toxicity, bioactivity, low wear rate, low to moderate incidence of artifacts during imaging, tensile strength and modulus that are comparable to those of cortical bone, high fatigue strength/long fatigue life, minimal or no negative impact on radiotherapy (RT) planning and delivery, and high capability for fusion to the contiguous bone. The shortcomings of Ti6Al4V alloy for these applications with respect to these desirable properties are well recognized, opening the field for an investigation about novel biomaterials that could replace the current gold standard.

View Article and Find Full Text PDF

Objective: To investigate the effect of Polyetheretherketone (PEEK) rod semi-rigid pedicle screw fixation system in lumbar spine non-fusion surgery.

Methods: A total of 74 patients with tow-level lumbar degenerative diseases who underwent surgery from March 2017 to December 2019 were divided into PEEK rod group and titanium rod group. In the PEEK rod group, there were 34 patients, including 13 males and 21 females, aged from 51 to 79 years old with an average of (62.

View Article and Find Full Text PDF
Article Synopsis
  • - The study investigates the use of Polyetheretherketone (PEEK) rods versus titanium rods in long-segment lumbar spine instrumentation, focusing on their potential to prevent adjacent segment degeneration.
  • - Researchers built finite element models to assess how different fixation methods (PEEK vs. titanium) affect biomechanical stress on vertebral components under a simulated load.
  • - Results indicate that PEEK rods may reduce stress on adjacent vertebral segments compared to titanium rods, suggesting PEEK could be more effective in preventing degeneration in spinal surgeries.
View Article and Find Full Text PDF

Objective: To investigate the correlation between magnetic resonance imaging-based vertebral bone quality (VBQ) score and screw loosening after dynamic pedicle screw fixation with polyetheretherketone (PEEK) rods, and evaluate its predictive value.

Methods: A retrospective analysis was conducted on the patients who underwent dynamic pedicle screw fixation with PEEK rods from March 2017 to June 2022. Data on age, sex, body mass index, hypertension, diabetes, hyperlipidemia history, long-term smoking, alcohol consumption, VBQ score, L1-4 average Hounsfield unit (HU) value, surgical fixation length, and the lowest instrumented vertebra were collected.

View Article and Find Full Text PDF

Retrieval analysis of PEEK rods pedicle screw system: three cases analysis.

BMC Musculoskelet Disord

June 2024

Orthopedic Department, 960 Hospital of People's Liberation Army, NO.25 Shifan Road, Jinan, Shandong, 250031, China.

Purpose: To analyze the characteristics of PEEK rods retrieved in vivo, specifically their wear and deformation, biodegradability, histocompatibility, and mechanical properties.

Method: Six PEEK rods were retrieved from revision surgeries along with periprosthetic tissue. The retrieved PEEK rods were evaluated for surface damage and internal changes using Micro-CT, while light and electron microscopy were utilized to determine any histological changes in periprosthetic tissues.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!