Background And Purpose: In the next few years the number of facilities providing ion beam therapy with scanning beams will increase. An auditing process based on an end-to-end test (including CT imaging, planning and dose delivery) could help new ion therapy centres to validate their entire logistic chain of radiation delivery. An end-to-end procedure was designed and tested in both scanned proton and carbon ion beams, which may also serve as a dosimetric credentialing procedure for clinical trials in the future. The developed procedure is focused only on physical dose delivery and the validation of the biological dose is out of scope of the current work.

Materials And Methods: The audit procedure was based on a homogeneous phantom that mimics the dimension of a head (20 × 20 × 21 cm(3)). The phantom can be loaded either with an ionisation chamber or 20 alanine dosimeters plus 2 radiochromic EBT films. Dose verification aimed at measuring a dose of 10Gy homogeneously delivered to a virtual-target volume of 8 × 8 × 12 cm(3). In order to interpret the readout of the irradiated alanine dosimeters additional Monte Carlo simulations were performed to calculate the energy dependent detector response of the particle fluence in the alanine detector. A pilot run was performed with protons and carbon ions at the Heidelberg Ion Therapy facility (HIT).

Results: The mean difference of the absolute physical dose measured with the alanine dosimeters compared with the expected dose from the treatment planning system was -2.4 ± 0.9% (1σ) for protons and -2.2 ± 1.1% (1σ) for carbon ions. The measurements performed with the ionisation chamber indicate this slight underdosage with a dose difference of -1.7% for protons and -1.0% for carbon ions. The profiles measured by radiochromic films showed an acceptable homogeneity of about 3%.

Conclusions: Alanine dosimeters are suitable detectors for dosimetry audits in ion beam therapy and the presented end-to-end test is feasible. If further studies show similar results, this dosimetric audit could be implemented as a credentialing procedure for clinical proton and carbon beam delivery.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.radonc.2013.04.029DOI Listing

Publication Analysis

Top Keywords

alanine dosimeters
20
ion beam
12
beam therapy
12
carbon ions
12
end-to-end test
8
dose
8
dose delivery
8
ion therapy
8
proton carbon
8
credentialing procedure
8

Similar Publications

ESR measurement of carbonated hydroxyapatite for dosemeter.

Radiat Prot Dosimetry

November 2024

Department of Advanced Quantum Beam Technology, Takasaki Institute for Advanced Quantum Science, National Institutes for Quantum Science and Technology (QST), 1233 Watanuki-cho, Takasaki, Gunma 370-1292, Japan.

We have examined a dosimetry characteristic of carbonate hydroxyapatite (CO3HAp), which is a dental bone graft material. The purpose of this work is to investigate the reproducibility and stability of radiation-induced radicals on CO3HAp samples and assess the feasibility of using these materials as dosemeters. CO3HAp samples were exposed to gamma rays with dose range from 10 to 10 000 Gy.

View Article and Find Full Text PDF

Alanine dosemeters are widely used as solid dosemeters; however, evaluating doses in the mGy order of radiation is difficult because of their lower sensitivity for low doses. We applied the electron spin resonance dosimetry technique to investigate whether hydroxyapatite, the main component of teeth, is a suitable material for a new dosemeter to detect radiation from mGy to Gy. Commercially available synthetic hydroxyapatite possesses the essential characteristics required for a dosemeter material and has been demonstrated to accurately measure doses in the mGy to Gy range.

View Article and Find Full Text PDF

Precise estimation of individual radiation dose utilizing biomaterials (fingernail, bone, and tooth) is very challenging due to their complex sample processing. Despite, tooth enamel, the most mineralized tissue of tooth is used for this purpose due to its high radiation sensitivity and ability to produce radiation induced long lived CO radicals. However, human teeth are not always available, and invasive nature of sample collection adds to the complexity making dose estimation difficult.

View Article and Find Full Text PDF

Background: The newly available ZAP-X stereotactic radiosurgical system is designed for the treatment of intracranial lesions, with several unique features that include a self-shielding, gyroscopic gantry, wheel collimation, non-orthogonal kV imaging, short source-axis distance, and low-energy megavoltage beam. Systematic characterization of its radiation as well as other properties is imperative to ensure its safe and effective clinical application.

Purpose: To accurately determine the radiation output of the ZAP-X with a special focus on the smaller diameter cones and an aim to provide useful recommendations on quantification of small field dosimetry.

View Article and Find Full Text PDF

The successful implementation of FLASH radiotherapy in clinical settings, with typical dose rates >40 Gy s, requires accurate real-time dosimetry.Silicon carbide (SiC) p-n diode dosimeters designed for the stringent requirements of FLASH radiotherapy have been fabricated and characterized in an ultra-high pulse dose rate electron beam. The circular SiC PiN diodes were fabricated at IMB-CNM (CSIC) in 3m epitaxial 4H-SiC.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!