Validation of quantitative bound- and pore-water imaging in cortical bone.

Magn Reson Med

Department of Biomedical Engineering, Vanderbilt University, Nashville, Tennessee, USA; Vanderbilt University Institute of Imaging Science, Vanderbilt University, Nashville, Tennessee, USA.

Published: June 2014

Purpose: To implement and validate a previously proposed ultra-short echo time method for measuring collagen-bound- and pore-water concentrations in bone based on their T2 differences.

Methods: Clinically compatible ultra-short echo time image sequences for quantitative T2 -based bound and pore-water imaging in bone were implemented and validated on a 3T human scanner and a 4.7T small bore system. Bound- and pore-water images were generating using T2 -selective adiabatic pulses. In both cases, the magnetization preparation was integrated into a three-dimensional ultra-short echo time acquisition, with 16 radial spokes acquired per preparation. Images were acquired from human cadaveric femoral mid-shafts from which isolated bone samples were subsequently extracted for nonimaging analysis using T2 spectroscopic measurements.

Results: A strong correlation was found between imaging-derived concentrations of bound and pore water and those determined from the isolated bone samples.

Conclusions: These studies demonstrate the translation of the previously developed approaches for distinguishing bound and pore water from human cortical bone using practical human MRI constraints of gradient performance and radiofrequency power deposition.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3899101PMC
http://dx.doi.org/10.1002/mrm.24870DOI Listing

Publication Analysis

Top Keywords

ultra-short echo
12
echo time
12
bound- pore-water
8
pore-water imaging
8
cortical bone
8
isolated bone
8
bound pore
8
pore water
8
bone
6
validation quantitative
4

Similar Publications

Utilizing a multi-task deep learning framework, this study generated synthetic CT (sCT) images from a limited dataset of Ultrashort echo time (UTE) MRI for transcranial focused ultrasound (tFUS) planning. A 3D Transformer U-Net was employed to produce sCT images that closely replicated actual CT scans, demonstrated by an average Dice coefficient of 0.868 for morphological accuracy.

View Article and Find Full Text PDF

Aims: The aim of this study was to develop an ultra-short echo time 3D magnetic resonance imaging (MRI) method for imaging subacute myocardial infarction (MI) quantitatively and in an accelerated way. Here, we present novel 3D T- and T -weighted Multi-Band SWeep Imaging with Fourier Transform and Compressed Sensing (MB-SWIFT-CS) imaging of subacute MI in mice hearts .

Methods And Results: Relaxation time-weighted and under-sampled 3D MB-SWIFT-CS MRI were tested with manganese chloride (MnCl) phantom and mice MI model.

View Article and Find Full Text PDF
Article Synopsis
  • MRI with radial scanning was used to create detailed images of rat lungs, both healthy and those with pulmonary hypertension.
  • The retrospective gating method enabled the analysis of lung function during different breathing phases, revealing decreased lung volumes and tidal volumes in the rats with pulmonary hypertension.
  • Ventilation maps indicated a significant reduction in airflow in the affected areas, showcasing the method's effectiveness in assessing lung pathology and its influence on respiratory function.
View Article and Find Full Text PDF

Differentiation of histological calcification classifications in breast cancer using ultrashort echo time and chemical shift-encoded imaging MRI.

Front Oncol

December 2024

Newcastle Magnetic Resonance Centre, Translational and Clinical Research Institute, Faculty of Medical Sciences, Newcastle University, Newcastle, United Kingdom.

Introduction: Ductal carcinoma (DCIS) accounts for 25% of newly diagnosed breast cancer cases with only 14%-53% developing into invasive ductal carcinoma (IDC), but currently overtreated due to inadequate accuracy of mammography. Subtypes of calcification, discernible from histology, has been suggested to have prognostic value in DCIS, while the lipid composition of saturated and unsaturated fatty acids may be altered in synthesis with potential sensitivity to the difference between DCIS and IDC. We therefore set out to examine calcification using ultra short echo time (UTE) MRI and lipid composition using chemical shift-encoded imaging (CSEI), as markers for histological calcification classification, in the initial step towards application.

View Article and Find Full Text PDF

Magnetic resonance imaging (MRI) techniques that enhance the visualization of mineralized tissues (hereafter referred to as MT-MRI) are increasingly being incorporated into clinical practice, particularly in musculoskeletal imaging. These techniques aim to mimic the contrast provided by computed tomography (CT), while taking advantage of MRI's superior soft tissue contrast and lack of ionizing radiation. However, the variety of MT-MRI techniques, including three-dimensional gradient-echo, ultra-short and zero-echo time, susceptibility-weighted imaging, and artificial intelligence-generated synthetic CT, each offer different technical characteristics, advantages, and limitations.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!