Background: There is no scientific literature that examines radiofrequency (RF) interference from Smart Meters with cardiac implantable electronic devices (CIEDs). The objective of this in vitro study was to assess any potential interference with Medtronic CIEDs (Medtronic Inc., Minneapolis, MN, USA).

Methods: In the Quebec testing, five models of Medtronic CIEDs were placed in an acrylic cylinder filled with a saline solution and faced a Landis+Gyr Smart Meter or Router (Landis+Gyr AG, Zug, Switzerland). The distance between CIEDs and the meter casing or router antenna was 10 cm. The Meter's normal operating conditions were modified to artificially set the number of impulsions at an abnormally high level (one, two, and three impulses per second). Each scenario was repeated one to five times, for 1 minute each. In the U.S. testing, 6 cm and 15 cm (∼2.25' and ∼6') separated the six models of Medtronic CIEDs from the Schlumberger Smart Meter (Itron Inc., Liberty Lake, WA, USA), which generally sent out a 96-bit Standard Consumption Message over 3 seconds. The transmission varied in frequencies along with the interval between cycles.

Results: A total of 6,966 RF transmissions were completed during the 34 tests conducted in Quebec. In the United States, the CIED was exposed to the meter for 10 minutes to provide a minimum of 200 completed RF transmissions. No interference was observed in worst-case scenarios (testing of meters and CIEDs at their performance limits).

Conclusions: Landis+Gyr Smart Meters/Routers and Schlumberger Smart Meters do not interfere with the functioning of the Medtronic CIEDs tested, when placed, respectively, 10 cm and 6 cm and 15 cm apart.

Download full-text PDF

Source
http://dx.doi.org/10.1111/pace.12225DOI Listing

Publication Analysis

Top Keywords

medtronic cieds
16
smart meters
12
models medtronic
8
landis+gyr smart
8
smart meter
8
schlumberger smart
8
cieds
7
smart
6
medtronic
5
meters routers
4

Similar Publications

[Remote monitoring for patients with cardiac implantable electronic devices and heart failure].

Herzschrittmacherther Elektrophysiol

January 2025

Hannover Herzrhythmus Centrum, Klinik für Kardiologie und Angiologie, Medizinische Hochschule Hannover, Carl-Neuberg-Str. 1, 30625, Hannover, Deutschland.

The digitalization in healthcare facilitates continuous monitoring of relevant medical parameters through internal and external sensors. For patients with heart failure and cardiac implantable electronic devices (CIEDs), telemedicine has the potential to improve patient care and reduce use of healthcare resources. Remote monitoring shortens the response time to a clinical event, reduces inappropriate shocks, and increases patient satisfaction.

View Article and Find Full Text PDF

Cardiac implantable electronic devices (CIEDs) generate substantial data, often stored in image or PDF formats. Remote monitoring, now an integral component of patient care, places considerable administrative burdens on clinicians and staff, in large part due to the challenge of integrating these data seamlessly into electronic health records. Since 2006, the Heart Rhythm Society, in collaboration with the CIED industry, has led an initiative to establish a unified standard nomenclature.

View Article and Find Full Text PDF

Aims: Utilization of transvenous lead extraction/removal (TLE) for the management of cardiac implantable electronic device (CIED)-associated infective endocarditis (IE) remains low. The aim of this study was to examine the impact of hospital TLE procedural volume on TLE utilization and outcomes for patients with CIED-associated IE.

Methods And Results: Using the Nationwide Readmissions Database, we evaluated 21 545 admissions for patients (mean age 70 years, 39% female) with CIEDs hospitalized with IE at TLE centres.

View Article and Find Full Text PDF

Racial Differences in Device-Detected Incident Atrial Fibrillation.

JACC Clin Electrophysiol

November 2024

Division of Cardiology, Northwestern University, Feinberg School of Medicine, Chicago, Illinois, USA. Electronic address:

Background: Prior research suggests Black individuals have a lower risk of atrial fibrillation (AF) than White individuals, but this may be related to the underdetection of AF. Whether this trend persists using highly sensitive methods of AF diagnosis has not been well-studied.

Objectives: The objective of this study was to use cardiac implantable electronic devices (CIEDs) capable of AF diagnosis to compare AF incidence between Black and White individuals.

View Article and Find Full Text PDF

Unlabelled: New cardiac implantable electronic devices (CIEDs), such as leadless pacemakers and subcutaneous implantable cardioverter defibrillators (S-ICDs), are being used in patients with adult congenital heart disease. The selection of CIEDs often requires careful consideration due to technical challenges posed by a unique heart structure. A 27-year-old man following a surgical tetralogy of Fallot (TOF) repair developed non-sustained ventricular tachycardia, sick sinus syndrome, and complete atrioventricular block.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!