Influence of Off-resonance in myocardial T1-mapping using SSFP based MOLLI method.

J Cardiovasc Magn Reson

National Heart, Lung, and Blood Institute, National Institutes of Health, DHHS, Bethesda, MD 20892, USA.

Published: July 2013

Background: Myocardial T1-mapping methods such as MOLLI use SSFP readout and are prone to frequency-dependent error in T1-measurement. A significant error in T1 may result at relatively small off-resonance frequencies that are well within the region without banding artifacts.

Methods: The sensitivity of T1-estimates based on the SSFP based MOLLI sequence to errors in center frequency are calculated by means of a Bloch simulation and validated by phantom measurements. Typical off-resonance errors following local cardiac shimming are determined by field mapping at both 1.5 and 3.0T. In vivo examples demonstrate the artifactual appearance of T1-maps in the presence of off-resonance variation.

Results: Off-resonance varied 61.8 ± 15.5 Hz (mean ± SD, n = 18) across the heart at 1.5T and 125.0 ± 40.6 Hz (mean ± SD, n = 18) at 3.0T. For T1 = 1000 ms, the variation in T1 due to off-resonance variation was approximately 20 ms at 62 Hz, and > 50 ms at 125 Hz.

Conclusions: Regional variations due to the inability to completely shim the B0-field variation around the heart appear as regional variation in T1, which is artifactual.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3733653PMC
http://dx.doi.org/10.1186/1532-429X-15-63DOI Listing

Publication Analysis

Top Keywords

myocardial t1-mapping
8
ssfp based
8
based molli
8
off-resonance
5
influence off-resonance
4
off-resonance myocardial
4
t1-mapping ssfp
4
molli method
4
method background
4
background myocardial
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!