Objectives: Since blood flow impairment by aortic valve prosthesis is characteristically dynamic, this dynamic component is best and thoroughly appreciated by exercise Doppler echocardiography. We sought to determine the haemodynamics of a new pericardial aortic bioprosthesis [Trifecta™-aortic valve bioprosthesis (T-AVB), St Jude Medical, MN, USA] at rest and during exercise and a 10-min recovery period in comparison with alternative aortic valve prostheses, e.g. Ross operation (RO), stentless aortic valve [Medtronic freestyle-aortic valve bioprosthesis (MF-AVB)] and a healthy control group (CO).
Methods: Haemodynamics at rest and during supine exercise stress testing and a 10-min recovery period were evaluated in 32 patients (mean age: 70.8 ± 6.7 years) with T-AVB (mean follow-up: 5 ± 2 months), 49 with RO (mean age: 43.5 ± 13.7 years), 39 with an MF-AVB (mean age: 64.6 ± 9.4 years) and 26 healthy patients (mean age: 39 ± 9 years). Measurements included mean outflow tract gradient (δp mean, mmHg), effective orifice area index (EOAI, cm(2)/m(2)) and valvular resistance (vR, dyn s cm(-5)).
Results: Mean body surface area for T-AVB was 1.93 ± 0.24 m(2) (median 1.97 m(2)). Mean δp mean at rest was 7.2 ± 3.4 mmHg, mean EOAI 0.86 ± 0.23 cm(2)/m(2) and mean vR 50.7 ± 23.2 dyn s cm(-5). Supine stress testing did increase the mean EOAI to 0.98 ± 0.27 cm(2)/m(2), the mean vR to 62.6 ± 25.3 dyn s cm(-5) and the mean δp mean to 10.21 ± 4.7 mmHg, respectively (P < 0.05 for all comparisons). During the post-exercise recovery period, δp mean, EOAI and vR showed a prompt normalization within 5 min of cessation of exercise. At all the three measurement points, T-AVB and MF-AVB revealed low gradients, satisfactory EOAI and low vR. Compared with the RO and a healthy control group, both groups showed significantly inferior performance throughout the exercise and post-exercise study protocol (P < 0.05). In comparison with T-AVB, patients with an MF-AVB only showed significant inferior performance throughout series with respect to a higher vR, indicating a smaller increase in the EOAI during exercise. During the 10-min post-exercise period, T-AVB recovered significantly earlier than MF-AVB.
Conclusions: When comparing two different types of aortic valve bioprostheses with a gold standard group (RO) and a healthy population, both aortic valve bioprostheses perform inferior but reveal promising haemodynamics during exercise. During post-exercise haemodynamic recovery, only the T-AVB revealed a nearly physiological recovery pattern compared with the RO and a healthy control group.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1093/ejcts/ezt367 | DOI Listing |
J Soc Cardiovasc Angiogr Interv
December 2024
Cardiovascular Institute, Detroit Medical Center, DMC Heart Hospital, Detroit, Michigan.
Transcatheter aortic valve repair (TAVR) presents a minimally invasive alternative to traditional surgical valve replacement, albeit not without its own set of complications. A rare complication is the infolding of the self-expanding valve, which can precipitate cardiac arrest. The estimated incidence rate of this complication stands at 1.
View Article and Find Full Text PDFCatheter Cardiovasc Interv
January 2025
Department of Medicine, Michael E DeBakey Veteran's Affairs Medical Center, Houston, Texas, USA.
Transcatheter aortic valve replacement (TAVR) has become a viable treatment option for patients with severe aortic stenosis among all risk subsets. As TAVR use becomes more prevalent and patients live longer with their transcatheter valve, an increasing number of these patients can be expected to present with ACS. Overall, there is a paucity of high-quality data detailing incidence, pathophysiology, and management of ACS in this subset.
View Article and Find Full Text PDFZhonghua Xin Xue Guan Bing Za Zhi
January 2025
Cardiology Department, Second Affiliated Hospital of Dalian Medical University, Dalian116021, China.
To assess the effectiveness of transcatheter aortic valve replacement (TAVR) on electrocardiographic remodeling in patients with severe aortic stenosis (AS), and identify its influencing factors. A cohort study was conducted on patients with a confirmed diagnosis of severe AS who successfully underwent TAVR at the Second Affiliated Hospital of Dalian Medical University between June 2018 and March 2023. Data, including standard 15-lead electrocardiograms and echocardiograms, were collected before the operation, 1 week after the operation, and 3 months after the operation.
View Article and Find Full Text PDFRen Fail
December 2025
Center for Cardiac Intensive Care, Beijing Anzhen Hospital, Capital Medical University, Beijing Institute of Heart, Lung and Blood Vessel Diseases, Beijing, China.
Background: The incidence of acute kidney injury (AKI) increases after surgical aortic valve replacement (SAVR). This study aimed to characterize the risk factors of AKI after SAVR.
Methods And Results: We conducted a retrospective registry study based on data from 299 consecutive patients undergoing SAVR.
Background: Surgical aortic valve replacement (SAVR) is the commonly used approach for aortic valve replacement (AVR) in patients with aortic stenosis at low or intermediate surgical risk. However, transcatheter aortic valve replacement (TAVR) has emerged as an alternative to SAVR for AVR. This meta-analysis aims to assess the comparative efficacy and safety of TAVR versus SAVR in low-to-intermediate surgical risk patients by analyzing temporal trends in the outcomes of TAVR and SAVR at various follow-up intervals, providing a more detailed understanding.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!