  • Olympus developed a new colonoscope prototype (XCF-Q180AY2L) featuring responsive insertion technology aimed at improving ease of colonoscope insertion during procedures.
  • A study involving 288 patients compared this prototype to conventional colonoscopes, measuring cecal intubation rates and times, with results indicating that the prototype took less time to reach the cecum and required fewer maneuvers.
  • The findings showed that both groups achieved 100% cecal intubation, but the prototype significantly reduced the subjective difficulty and time needed for the procedure, suggesting it may be a more effective tool for colonoscopy.

Article Abstract

Background And Objectives: recently, Olympus developed a new prototype (XCF-Q180AY2L) with responsive insertion technology (RIT), which besides the still known variable stiffness technology, included a passive bending section and a new high force insertion tube. Our objective was to investigate whether the use of this prototype could ease colonoscope insertion, increasing the cecal intubation rate, and/or shortening the cecal intubation time.

Material And Methods: the study was designed as a prospective observational study in 305 consecutive patients from a colo-rectal cancer screening program. We compared colonoscopies performed with conventional colonoscopes (CFH180AL/CFQ160L) with those performed with the prototype XCF-Q180AY2L. End points were mean cecal and terminal ileal intubation times, cecum intubation rate, and need for specific maneuvers. All colonoscopies were performedunder sedation with intravenous propofol. Finally, 288 patients were included.

Results: no complications were observed. Complete cecal intubation rate was 100 % in both groups. The ileum could be reached in 98.95 % of cases. Mean time required to reach the cecum was shorter in the prototype endoscope group (4.31 min, SD 2.63 min) than in the conventional endoscope group (4.66 min, SD 2.52 min) (p < 0.05). Compared with the standard colonoscope group, we observed in the prototype group less subjective sensation of difficultyin the passage of the sigma (p < 0.01), fewer maneuvers when it proved necessary to straighten the scope (p < 0.01), and less frequent need to modify the stiffness of the endoscope (p < 0.05).

Conclusion: we concluded that the prototype endoscope (XCFQ180AY2L) facilitated colonoscope insertion, requiring slightly less time to reach the cecum than a standard colonoscope.

Download full-text PDF

Source
http://dx.doi.org/10.4321/s1130-01082013000400005DOI Listing

Publication Analysis

Top Keywords

cecal intubation
12
intubation rate
12
responsive insertion
8
insertion technology
8
technology rit
8
prototype xcf-q180ay2l
8
colonoscope insertion
8
reach cecum
8
prototype endoscope
8
endoscope group
8

Similar Publications

British Society of Gastroenterology National Evaluation of Colonoscopy Quality: Findings from the National Endoscopy Database.

Gastrointest Endosc

January 2025

Population Health Sciences Institute, Newcastle University Centre for Cancer, Newcastle University, Newcastle-upon-Tyne, UK, NE2 4AX; North Tees and Hartlepool NHS Foundation Trust, TS19 8PE.

Background And Aims: Analysis of national colonoscopy quality using automatically uploaded data from a national database, including exploring performance variation.

Methods: Data on all colonoscopies performed in the UK 01/03/2019-29/02/2020 and recorded in the National Endoscopy Database were analysed. Unadjusted key performance indicators were calculated and proportions of endoscopists achieving national standards were determined.

View Article and Find Full Text PDF

Introduction: Controversy remains regarding transparent cap-assisted technique improves adenoma detection rate (ADR) in colonoscopy. We aimed to investigate the effect of transparent cap-assisted colonoscopy on ADR and other colonoscopy performance.

Methods: We performed sub-analysis of an international, multicenter, open-label database containing colonoscopy data from 11 centers in 4 Asian countries/regions on patients who underwent colonoscopy.

View Article and Find Full Text PDF

Improving colonoscopy quality in the national VA healthcare system.

Contemp Clin Trials

December 2024

San Francisco VA Medical Center, United States of America; Department of Medicine, University of California, San Francisco, CA, United States of America; Measurement Science Quality Enhancement Research Initiative, San Francisco VA Healthcare System, United States of America. Electronic address:

Background: Colorectal cancer (CRC) prevention is a Veterans Affairs (VA) priority. Colonoscopy quality, especially adenoma detection rate (ADR), is critical for effective screening. Our research indicates considerable variation in ADR among VA providers.

View Article and Find Full Text PDF

A longer cecal intubation time (CIT) occurs during colonoscopy under difficult insertion conditions, which may hinder meticulous mucosal observation. However, whether a longer CIT has detrimental effects on the detection of adenomas remains unclear. We evaluated the effects of CIT on the detection and miss rates of colorectal neoplasms in asymptomatic participants.

View Article and Find Full Text PDF

By using optimal insertion techniques with water infusion and dynamic position changes, pain during colonoscopy is greatly reduced and the procedures can usually be performed without sedation. We investigated whether the excellent results with water-aided colonoscopy reported by experts are reproducible in daily practice in a regional hospital. During the year 2023, 500 consecutive outpatients 50-75 years old presenting for colorectal cancer screening and surveillance could choose between unsedated or on-demand minimally sedated colonoscopy, moderate sedation with midazolam, or deep sedation with propofol.

View Article and Find Full Text PDF