Postoperative displacement of deep brain stimulation electrodes related to lead-anchoring technique.

Neurosurgery

Departments of †Neurology; ‡Neurosurgery; and §Psychiatry, Academic Medical Center, University of Amsterdam, Amsterdam, the Netherlands; ¶Department of Neurology, University Medical Centre Groningen, Groningen, the Netherlands.

Published: October 2013

Background: Displacement of deep brain stimulation (DBS) electrodes may occur after surgery, especially due to large subdural air collections, but other factors might contribute.

Objective: To investigate factors potentially contributing to postoperative electrode displacement, in particular, different lead-anchoring techniques.

Methods: We retrospectively analyzed 55 patients (106 electrodes) with Parkinson disease, dystonia, tremor, and obsessive-compulsive disorder in whom early postoperative and long-term follow-up computed tomography (CT) was performed. Electrodes were anchored with a titanium microplate or with a commercially available plastic cap system. Two independent examiners determined the stereotactic coordinates of the deepest DBS contact on early postoperative and long-term follow-up CT. The influence of age, surgery duration, subdural air volume, use of microrecordings, fixation method, follow-up time, and side operated on first was assessed.

Results: Subdural air collections measured on average 4.3 ± 6.2 cm. Three-dimensional (3-D) electrode displacement and displacement in the X, Y, and Z axes significantly correlated only with the anchoring method, with larger displacement for microplate-anchored electrodes. The average 3-D displacement for microplate-anchored electrodes was 2.3 ± 2.0 mm vs 1.5 ± 0.6 mm for electrodes anchored with the plastic cap (P = .030). Fifty percent of the microplate-anchored electrodes showed 2-mm or greater (potentially relevant) 3-D displacement vs only 25% of the plastic cap-anchored electrodes (P < .01).

Conclusion: The commercially available plastic cap system is more efficient in preventing postoperative DBS electrode displacement than titanium microplates. A reliability analysis of the electrode fixation is warranted when alternative anchoring methods are used.

Download full-text PDF

Source
http://dx.doi.org/10.1227/NEU.0000000000000079DOI Listing

Publication Analysis

Top Keywords

subdural air
12
electrode displacement
12
plastic cap
12
microplate-anchored electrodes
12
electrodes
9
displacement deep
8
deep brain
8
brain stimulation
8
displacement
8
air collections
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!