In this prospective study, cardiac output was measured in 38 intensive care unit patients before and after a fluid challenge, using both pulse contour analysis (Nexfin(®); BMEYE, Amsterdam, the Netherlands) and transthoracic echocardiography. The ability of the Nexfin device to detect significant changes in the velocity-time integral was evaluated. The pulse wave could not be detected by the Nexfin device in five patients (13%), leaving 33 patients for analysis. The Nexfin device adequately tracked changes in the velocity-time integral in 20 (61%) patients. Using a cut-off of a 10% increase in cardiac output estimated by the Nexfin or by echocardiography, the sensitivity of the Nexfin device to detect a response to fluid challenge was 47%, with specificity 81% and accuracy 64%. The percentage error between the Nexfin and echocardiography was 448%; lower limit of agreement -48% (95% CI -62 to -36%) and upper limit of agreement, 32% (95% CI 20-45%). We conclude that the Nexfin device does not adequately track changes in cardiac output in critically ill patients.