Purpose: The use of small photon fields is now an established practice in stereotactic radiosurgery and radiotherapy. However, due to a lack of lateral electron equilibrium and high dose gradients, it is difficult to accurately measure the dosimetric quantities required for the commissioning of such systems. Moreover, there is still no metrological dosimetric reference for this kind of beam today. In this context, the first objective of this work was to determine and to compare small fields output factors (OF) measured with different types of active detectors and passive dosimeters for three types of facilities: a CyberKnife(®) system, a dedicated medical linear accelerator (Novalis) equipped with m3 microMLC and circular cones, and an adaptive medical linear accelerator (Clinac 2100) equipped with an additional m3 microMLC. The second one was to determine the kQclin,Qmsr (fclin,fmsr) correction factors introduced in a recently proposed small field dosimetry formalism for different active detectors.

Methods: Small field sizes were defined either by microMLC down to 6 × 6 mm(2) or by circular cones down to 4 mm in diameter. OF measurements were performed with several commercially available active detectors dedicated to measurements in small fields (high resolution diodes: IBA SFD, Sun Nuclear EDGE, PTW 60016, PTW 60017; ionizing chambers: PTW 31014 PinPoint chamber, PTW 31018 microLion liquid chamber, and PTW 60003 natural diamond). Two types of passive dosimeters were used: LiF microcubes and EBT2 radiochromic films.

Results: Significant differences between the results obtained by several dosimetric systems were observed, particularly for the smallest field size for which the difference in the measured OF reaches more than 20%. For passive dosimeters, an excellent agreement was observed (better than 2%) between EBT2 and LiF microcubes for all OF measurements. Moreover, it has been shown that these passive dosimeters do not require correction factors and can then be used as reference dosimeters. Correction factors for the active detectors have then been determined from the mean experimental OF measured by the passive dosimeters.

Conclusions: Four sets of correction factors needed to apply the new small field dosimetry formalism are provided for several active detectors. A protocol for small photon beams OF determination based on passive dosimeters measurements has been recently proposed to French radiotherapy treatment centers.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.4811139DOI Listing

Publication Analysis

Top Keywords

correction factors
20
passive dosimeters
20
active detectors
16
small fields
12
circular cones
12
small field
12
small
8
fields output
8
output factors
8
equipped micromlc
8

Similar Publications

Purpose: To study the clinical profile of patients with microspherophakia and the factors associated with poor vision following lensectomy surgery.

Methods: A multicenter, retrospective, cross-sectional analysis was conducted from January 2010 to June 2022 on patients diagnosed with microspherophakia.

Results: A total of 102 eyes from 51 patients were enrolled, of whom 24 (47.

View Article and Find Full Text PDF

Background: The vertical vein (VV) ligation during the total anomalous pulmonary venous return (TAPVR) correction is still controversial. Our study aimed to define the potential risk factors for VV persistence and their percutaneous occlusion.

Methods: The retrospective cohort study included 40 patients (26 males) with TAPVR treated at the tertiary referral center from 2005 to 2024.

View Article and Find Full Text PDF

Very preterm infants (VPIs) often experience extrauterine growth failure. Therefore, aggressive nutritional management of VPIs is recommended with the goal of achieving the postnatal growth of an equivalent fetus. However, VPIs frequently present postnatal length growth restriction at term-corrected age that remains lower than the standard weight and have greater fat mass and lower lean and bone mass than term-born infants.

View Article and Find Full Text PDF

Objective: Inflammatory factors play a crucial role in the onset and progression of heart failure. To further explore the causal relationship between inflammatory factors and heart failure, we employed bidirectional Mendelian randomization analysis to investigate the causal links between 91 inflammatory cytokines and heart failure.

Methods: We conducted our study using the bidirectional Mendelian randomization approach.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!