Introduction And Hypothesis: Pullout force of mesh from tissue is one of the important mechanical properties of an implanted mesh to repair pelvic organ prolapse (POP). The EndoFast Reliant™ system kit allows mesh attachment with soft-tissue fasteners. The aim of this study was to compare the pullout force that developed in mesh that was attached by EndoFast Reliant fasteners to mesh that was attached by trocar-based methods (tunnel, pocket) in a sheep model.

Methods: Six sheep underwent mesh attachment with three methods (EndoFast Reliant, tissue pocket, tissue tunnel), and each method was repeated five times in both thighs of the same sheep. The pullout force was measured at different time intervals from surgery: 0, 3, 7, 15, 30, and 45 days. Statistical analysis was performed by using the appropriate one-way analysis of variance (ANOVA) for each time interval and a general linear model for repeated measures using IBM® SPSS® software version 20.0.0.

Results: During the immediate postoperative period (0-3 days), pullout force was significantly higher with EndoFast Reliant than with tissue pocket or tissue tunnel. At day 7, this trend continues without statistical significance. Pullout force increased progressively until day 15, when the force caused the mesh to tear; it was similar in all three groups. ANOVA showed significant effect of time and study group.

Conclusions: The EndoFast Reliant system provides significantly stronger attachment in the immediate postoperative period (0-3 days) compared with trocar-based techniques, and this difference disappeared at day 15 postsurgery.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00192-013-2162-3DOI Listing

Publication Analysis

Top Keywords

pullout force
24
endofast reliant
16
mesh
8
mesh attachment
8
mesh attached
8
reliant tissue
8
tissue pocket
8
pocket tissue
8
tissue tunnel
8
anova time
8

Similar Publications

Objective: Combining oblique lumbar interbody fusion (OLIF) with posterior pedicle screw fixation (PPSF) has been proposed to reduce cage subsidence, especially in osteoporotic spines. Recently, anterolateral screw-rod fixation has gained interest as it allows direct pathology observation and avoids a posterior approach. However, controversies exist between anterolateral screw fixation systems and traditional PPSF due to variations in osteoporotic vertebral mineral density, screw fixation positions, and fixation methods (bicortical vs.

View Article and Find Full Text PDF

Atomistic Simulations of Mechanical Properties of Lignin.

Polymers (Basel)

December 2024

Department of Chemistry and Biochemistry, University of South Carolina, Columbia, SC 29208, USA.

The mechanical properties of lignin, an aromatic heteropolymer constituting 20-30% plant biomass, are important to the fabrication and processing of lignin-based sustainable polymeric materials. In this study, atomistic simulations are performed to provide microscopic insights into the mechanics of lignin. Representative samples of miscanthus, spruce, and birch lignin are studied.

View Article and Find Full Text PDF

Medial meniscus root tears (MMRTs) are serious injuries that disrupt knee biomechanics, often accelerating cartilage degeneration and osteoarthritis when left untreated. These injuries are increasingly recognized as a major cause of knee pain and functional limitations, particularly among middle-aged and older adults. This systematic review and meta-analysis aimed to evaluate the outcomes of conservative management compared to surgical intervention for MMRT, focusing on pain relief, functional recovery, and the progression of osteoarthritis.

View Article and Find Full Text PDF

Background: The fracture of an endodontic instrument within the root canal system can occur during root canal therapy, complicating thorough cleaning and shaping. Consequently, managing the broken fragment becomes crucial.

Methods: Eighty Nickel-titanium (NiTi) #20 K-files (Mani, Tochigi, Japan) were cut 8 mm from the tip, fixed into a corkboard, and classified into five groups (n = 14 each).

View Article and Find Full Text PDF

The new type of support disc-type anchor is an expanded body anchor with broad application prospects, and its load-bearing performance is significantly better than that of traditional anchors. However, there is a problem of premature shear damage in traditional support disc-type anchors. In order to solve this problem, this paper improves the traditional support disk anchor.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!