Background: Functional hemodynamic indicators (systolic pressure variation [SPV and SPV%] and pulse pressure variation [PPV%]) are sensitive and specific indicators of fluid responsiveness. It was unknown if these indicators could be accurately measured directly from the bedside monitor.

Objective: Determine the accuracy of SPV, SPV%, and PPV% measurements by using a stop-cursor method compared with a digitized analog strip (gold standard).

Methods: A prospective observational study using a convenience sample of 30 adult patients in a medical-surgical intensive care unit who were receiving mechanical ventilation and had no spontaneous breaths during 3 sequential ventilator breaths and had an optimized arterial catheter. The peak and nadir arterial pressure values for a ventilator cycle were simultaneously obtained by using the stop-cursor method on the bedside monitor and a hardcopy strip. The indicators were averaged over 3 breaths, and the difference between methods was calculated.

Results: Data were analyzed from 29 patients (1 patient excluded) on assist control ventilation (mean [SD] for tidal volume, 7.5 [2] mL/kg; positive end-expiratory pressure, 7 [4] cm H2O). For SPV, the mean bias was 0.4 (SD, 0.9) mm Hg (95% limits of agreement [LOA], -1.4 to 2.2 mm Hg); for SPV%, 0.3 (SD, 0.9; 95% LOA, -1.5% to 2.1%); for PPV%, 1.0 (SD, 3.3; 95% LOA, -5.5% to 7.5%). In only 1 case (PPV%) was there disagreement on fluid response characterization.

Conclusions: Statistically significant small differences in SPV and SPV% were detected. The differences in SPV, SPV%, and PPV% were not clinically significant, suggesting that functional hemodynamic indicators can be obtained accurately with the stop-cursor method.

Download full-text PDF

Source
http://dx.doi.org/10.4037/ajcc2013295DOI Listing

Publication Analysis

Top Keywords

stop-cursor method
16
pressure variation
12
spv spv%
12
pulse pressure
8
functional hemodynamic
8
hemodynamic indicators
8
indicators accurately
8
spv% ppv%
8
95% loa
8
differences spv
8

Similar Publications

Background: Functional hemodynamic indicators (systolic pressure variation [SPV and SPV%] and pulse pressure variation [PPV%]) are sensitive and specific indicators of fluid responsiveness. It was unknown if these indicators could be accurately measured directly from the bedside monitor.

Objective: Determine the accuracy of SPV, SPV%, and PPV% measurements by using a stop-cursor method compared with a digitized analog strip (gold standard).

View Article and Find Full Text PDF

Background: Pulmonary artery waveforms fluctuate because of changes in intrathoracic pressure caused by respirations. Monitoring system algorithms determine digital displays of pressure measurements on the basis of recognition, analysis, and comparison of consecutive waveforms.

Objective: To compare three methods of measuring pulmonary artery pressure during mechanical ventilation and spontaneous breathing in cardiac surgery patients with stable hemodynamics.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!