Background And Study Aims: MAGNAMOSIS forms a compression anastomosis using self-assembling magnetic rings that can be delivered via flexible endoscopy. The system has proven to be effective in full-thickness porcine small-bowel anastomoses. The aim of this study was to show the feasibility of the MAGNAMOSIS system in hybrid endoscopic colorectal surgery and to compare magnetic and conventional stapled anastomoses.

Methods: A total of 16 swine weighing 35 - 50 kg were used following animal ethical committee approval. The first animal was an acute model to establish the feasibility of the procedure. The subsequent 15 animals were survival models, 10 of which underwent side-to-side anastomoses (SSA) and 5 of which underwent end-to-side (ESA) procedures. Time to patency, surveillance endoscopy, burst pressure, compression force, and histology were assessed. Histology was compared with conventional stapled anastomoses. Magnetic compression forces were measured in various anastomosis configurations.

Results: Colorectal anastomoses were performed in all cases using a hybrid NOTES technique. The mean operating time was 71 minutes. Mean time to completion of the anastomosis was similar between the SSA and ESA groups. Burst pressure at 10 days was greater than 95 mmHg in both groups. One complication occurred in the ESA group. Compression force among various configurations of the magnetic rings was significantly different (P < 0.05). Inflammation and fibrosis were similar between magnetic SSA and conventional stapled anastomoses.

Conclusion: MAGNAMOSIS was feasible in performing a hybrid NOTES colorectal anastomosis. It has the advantage over circular staplers of precise endoscopic delivery throughout the entire colon. SSA was reliable and effective. A minimum initial compression force of 4 N appears to be required for reliable magnetic anastomoses.

Download full-text PDF

Source
http://dx.doi.org/10.1055/s-0033-1344119DOI Listing

Publication Analysis

Top Keywords

conventional stapled
12
compression force
12
magnetic compression
8
compression anastomosis
8
colorectal surgery
8
magnetic rings
8
burst pressure
8
hybrid notes
8
compression
6
magnetic
6

Similar Publications

Background: Total laparoscopic hysterectomy (TLH) is nowadays the standard to treat benign and malignant disease occurring in the uterus, but the number of robotic-assisted surgeries is increasing worldwide. To facilitate the handling of sutures in a bi- and tri-dimensional plane, a new type of suture material has been developed, named barbed sutures, which are in use in different indications. In comparison to conventional suture materials, the barbs anchor the suture in the tissue, provide tissue approximation and prevent slippage without the need for knot tying.

View Article and Find Full Text PDF

Purpose: The double-flap technique (DFT) is an anti-reflux reconstruction procedure performed after proximal gastrectomy (PG), but its complexity and high incidence of anastomotic stenosis are problematic. We conducted this study to demonstrate the efficacy of robot-assisted DFT, with refinements, to address these issues.

Methods: Surgical outcomes were compared between the following procedures modified over time at our institution: conventional open DFT (group O, n = 16); early robotic DFT (group RE, n = 19), which follows the conventional open PG approach; and late robotic DFT (group RL, n = 21), which incorporates refinements to the early robotic DFT technique by exploiting more of the robotic capabilities available.

View Article and Find Full Text PDF

Suture-associated surgical site infection (SSI) causes bacterial pathogens to colonize on the suture surface that are highly resistant to antibiotic treatment. Conventional suture materials used in surgical practice are causing complications such as infection and chronic inflammation. Surgical suture materials with antibacterial coatings are widely used in surgical practice.

View Article and Find Full Text PDF

Background: In the last years, the Cor-Knot device has been increasingly used in heart valve surgery. Our aim was to investigate the incidence of valvular complications in patients who underwent valvular surgery using the Cor-Knot device in multicentric cohorts at one-year follow-up.

Methods: Three hundred and sixty-eight patient underwent heart valve repair or replacement surgery using automated titanium suture fasteners in four cardiothoracic surgery departments between September 2018 and January 2020.

View Article and Find Full Text PDF

Is stapled Kono-S anastomosis a protective factor against postoperative endoscopic recurrence in Crohn's disease? a single-center, retrospective cohort study.

J Gastrointest Surg

January 2025

Center for Inflammatory Bowel Diseases, Department of General Surgery, The Affiliated Jinling Hospital of Nanjing Medical University, Nanjing, China; Center for Inflammatory Bowel Diseases, Department of General Surgery, Jinling Hospital, School of Medicine, Nanjing University, Nanjing, China. Electronic address:

Background: Hand-sewn Kono-S anastomosis is safe and associated with a reduction in post-operative recurrence (POR) in Crohn's disease (CD). The study aims to investigate the advantages of stapled Kono-S in Crohn's patients with intestinal anastomosis.

Methods: Crohn's patients undergoing intestinal anastomosis were reviewed via a prospectively maintained database.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!