Aim: Perfusion Index (PI) is a quantifiable measurement of peripheral perfusion and may be a useful adjunct to the assessment of circulatory status in the newborn. (i) To assess reproducibility of PI and (ii) To determine whether there is a difference between simultaneously obtained limb measurements of PI in newborns <32 weeks GA in the transitional period.

Methods: Perfusion Index was measured in newborns <32 weeks during the first 48 h of life. To examine reproducibility, the pulse oximetry probe was replaced on the same limb consecutively by the same operator. Upper and lower limbs were then simultaneously evaluated over a 5-min period. Heart rate, blood pressure, birth weight, ventilation requirement, inotrope use, lactate, PCO2 and CRIB-II score were also recorded.

Results: Thirty infants were assessed. Intraclass correlation coefficient for reproducibility in the same limb was high (r value = 0.982 p < 0.001). Measurements obtained in the right upper limb were consistently higher than either lower limb. The median (IQR) PI for the entire cohort was 0.70 (0.29-1.35). No correlation existed between gestational age, birth weight, CRIB scores, systolic and diastolic blood pressure, mean blood pressure and median PI values.

Conclusion: Perfusion Index measurement is reproducible, and values are highest in the right upper limb. Wide differences between right upper and lower limb readings are most likely related to transitional circulatory changes.

Download full-text PDF

Source
http://dx.doi.org/10.1111/apa.12322DOI Listing

Publication Analysis

Top Keywords

perfusion preterm
4
preterm infant
4
infant aim
4
aim perfusion
4
perfusion quantifiable
4
quantifiable measurement
4
measurement peripheral
4
peripheral perfusion
4
perfusion adjunct
4
adjunct assessment
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!