Introduction: Proximal humerus fracture fixation can be difficult because of osteoporosis making it difficult to achieve stable implant anchorage in the weak bone stock even when using locking plates. This may cause implant failure requiring revision surgery. Cement augmentation has, in principle, been shown to improve stability. The aim of this study was to investigate whether augmentation of particular screws of a locking plate aimed at a region of low bone quality is effective in improving stability in a proximal humerus fracture model.

Materials And Methods: Twelve paired human humerus specimens were included. Quantitative computed tomography was performed to determine bone mineral density (BMD). Local bone quality in the direction of the six proximal screws of a standard locking plate (PHILOS, Synthes) was assessed using mechanical means (DensiProbe™). A three-part fracture model with a metaphyseal defect was simulated and fixed with the plate. Within each pair of humeri the two screws aimed at the region of the lowest bone quality according to the DensiProbe™ were augmented in a randomised manner. For augmentation, 0.5 ml of bone cement was injected in a screw with multiple outlets at its tip under fluoroscopic control. A cyclic varus-bending test with increasing upper load magnitude was performed until failure of the screw-bone fixation.

Results: The augmented group withstood significantly more load cycles. The correlation of BMD with load cycles until failure and BMD with paired difference in load cycles to failure showed that augmentation could compensate for a low BMD.

Discussion And Conclusion: The results demonstrate that augmentation of screws in locked plating in a proximal humerus fracture model is effective in improving primary stability in a cyclic varus-bending test. The augmentation of two particular screws aimed at a region of low bone quality within the humeral head was almost as effective as four screws with twice the amount of bone cement. Screw augmentation combined with a knowledge of the local bone quality could be more effective in enhancing the primary stability of a proximal humerus locking plate because the effect of augmentation can be exploited more effectively limiting it to the degree required.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.injury.2013.05.008DOI Listing

Publication Analysis

Top Keywords

proximal humerus
20
bone quality
20
humerus fracture
12
augmentation screws
12
locking plate
12
aimed region
12
load cycles
12
augmentation
9
bone
9
screw augmentation
8

Similar Publications

Background: Chondroblastoma is classified as a benign bone tumor. However, postoperative local recurrence remains a concern. We analyzed the factors contributing to chondroblastoma local recurrence and the clinical challenges associated with treating these patients.

View Article and Find Full Text PDF

Background: The treatment of patients who suffer a proximal humeral fracture (PHF) remains controversial. The purpose of this study was to find consensus among experts using a validated iterative process in the treatment of patients after a PHF.

Methods: The Neer Circle is an organization of shoulder experts recognized for their service to the American Shoulder and Elbow Surgeons.

View Article and Find Full Text PDF

Sexual dimorphism of the humerus bones in a French sample: comparison of several statistical models including machine learning models.

Int J Legal Med

January 2025

Faculté de Médecine, Institut Universitaire d'Anthropologie Médico-Légale, Université Côte d'Azur, 28 Avenue de Valombrose, Nice Cedex 2, 06107, France.

Sex estimation is an important part of skeletal analysis and forensic identification. Traditionally pelvic traits are utilized for accurate sex estimation. However, the long bones, especially humerus, have been proved to be as effective for determine the sex of the individual.

View Article and Find Full Text PDF

The aim of this study is to investigate whether not using the calcar screw in proximal humerus fractures affects functional and radiological outcomes. Thirty patients (21 females and 9 males) who presented with proximal humerus fractures and were treated with plate- screw fixation were evaluated. The patients were divided into two groups: group 1 included patients with the use of the calcar screw, and group 2 included patients without the calcar screw.

View Article and Find Full Text PDF

Fractures of the humerus are common on the midshaft of the bone, often causing injury to the nutrient artery. Successful fracture repair and healing requires preservation of the blood supply to the long bones which is conveyed through the nutrient foramina (NF). The topography of long bone NF varies in different populations.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?