Impact of glucose measurement processing delays on clinical accuracy and relevance.

J Diabetes Sci Technol

Abbott Diabetes Care Inc., 1360 South Loop Rd., Alameda, CA 94502, USA.

Published: May 2013

Background: In a hospital setting, glucose is often measured from venous blood in the clinical laboratory. However, laboratory glucose measurements are typically not available in real time. In practice, turn-around times for laboratory measurements can be minutes to hours. This analysis assesses the impact of turn-around time on the effective clinical accuracy of laboratory measurements.

Methods: Data obtained from an earlier study with 58 subjects with type 1 diabetes mellitus (T1DM) were used for this analysis. In the study, glucose measurements using a YSI glucose analyzer were obtained from venous blood samples every 15 min while the subjects were at the health care facility. To simulate delayed laboratory results, each YSI glucose value from a subject was paired with one from a later time point (from the same subject) separated by 15, 30, 45, and 60 min. To assess the clinical accuracy of a delayed YSI result relative to a real-time result, the percentage of YSI pairs that meet the International Organization for Standardization (ISO) 15197:2003(E) standard for glucose measurement accuracy (±15 mg/dl for blood glucose < 75 mg/dl, ±20% for blood glucose ≥ 75 mg/dl) was calculated.

Results: It was observed that delays of 15 min or more reduce clinical accuracy below the ISO 15197:2003(E) recommendation of 95%. The accuracy was less than 65% for delays of 60 min.

Conclusion: This analysis suggests that processing delays in glucose measurements reduce the clinical relevance of results in patients with T1DM and may similarly degrade the clinical value of measurements in other patient populations.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3869134PMC
http://dx.doi.org/10.1177/193229681300700311DOI Listing

Publication Analysis

Top Keywords

clinical accuracy
16
glucose measurements
12
glucose
9
glucose measurement
8
processing delays
8
venous blood
8
ysi glucose
8
iso 151972003e
8
blood glucose
8
reduce clinical
8

Similar Publications

Prospective validation study of a combined urine and plasma test for predicting high-grade prostate cancer in biopsy naïve men.

Scand J Urol

January 2025

Department of Urology, Odense University Hospital, Odense, Denmark; Academy of Geriatric Cancer Research (AgeCare), Odense University Hospital, Odense, Denmark; Department of Clinical Research, University of Southern Denmark, Odense, Denmark.

Objective: Early and accurate diagnosis of prostate cancer (PC) is crucial for effective treatment. Diagnosing  clinically insignificant cancers can lead to overdiagnosis and overtreatment, highlighting the importance of accurately selecting patients for further evaluation based on improved risk prediction tools. Novel biomarkers offer promise for enhancing this diagnostic process.

View Article and Find Full Text PDF

Background: Clear cell renal cell carcinoma (ccRCC) is the most common subtype of renal cell carcinoma (RCC). Due to the lack of symptoms until advanced stages, early diagnosis of ccRCC is challenging. Therefore, the identification of novel secreted biomarkers for the early detection of ccRCC is urgently needed.

View Article and Find Full Text PDF

Validation Status of Electronic Sphygmomanometers in China: A National Survey.

Hypertension

January 2025

Clinical Research Institute, Institute of Advanced Clinical Medicine, Peking University, Beijing, China (X.Z., W.X., Y.W.).

Background: Although the information on the validation status of electronic sphygmomanometer (ES) devices in use in health care institutions and households is much more clinically relevant than that of ES models available on the market, it remains insufficient.

Methods: A national survey was conducted across all administrative regions of mainland China to assess the validation status of ESs. Fifty-eight cities were selected with stratification by municipality, provincial capital, and other cities, and health care institutions and households in each city were chosen by convenience to identify ES devices in use according to the study protocol.

View Article and Find Full Text PDF

Background/purpose: Computer-assisted implant surgery (CAIS) is increasingly performed to reduce deviations in implant position. Dynamic CAIS or navigation systems provide instant display of implant drilling instruments and patient positions directly on the computer monitor. Augmented reality (AR) technology allows operators to visualize real-time information projected onto the lenses of AR glasses.

View Article and Find Full Text PDF

The shift towards virtual assessment in dental education represents a significant advancement over traditional evaluation methods, offering new opportunities for assessing both theoretical knowledge and practical skills. This review examines the use of virtual assessment tools within the context of dental education. A narrative literature review was conducted, analyzing studies published between 2000 and 2024, sourced from PubMed, the Cochrane Library, and Embase.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!