Background And Objectives: During navigated procedures a tracked pointing device is used to define target structures in the patient to visualize its position in a registered radiologic data set. When working with endoscopes in minimal invasive procedures, the target region is often difficult to reach and changing instruments is disturbing in a challenging, crucial moment of the procedure. We developed a device for touch less navigation during navigated endoscopic procedures.

Materials And Methods: A laser beam is delivered to the tip of a tracked endoscope angled to its axis. Thereby the position of the laser spot in the video-endoscopic images changes according to the distance between the tip of the endoscope and the target structure. A mathematical function is defined by a calibration process and is used to calculate the distance between the tip of the endoscope and the target. The tracked tip of the endoscope and the calculated distance is used to visualize the laser spot in the registered radiologic data set.

Results: In comparison to the tracked instrument, the touch less target definition with the laser spot yielded in an over and above error of 0.12 mm. The overall application error in this experimental setup with a plastic head was 0.61 ± 0.97 mm (95% CI -1.3 to +2.5 mm).

Conclusion: Integrating a laser in an endoscope and then calculating the distance to a target structure by image processing of the video endoscopic images is accurate. This technology eliminates the need for tracked probes intraoperatively and therefore allows navigation to be integrated seamlessly in clinical routine. However, it is an additional chain link in the sequence of computer-assisted surgery thus influencing the application error.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3791553PMC
http://dx.doi.org/10.1002/lsm.22148DOI Listing

Publication Analysis

Top Keywords

laser spot
12
navigated procedures
8
registered radiologic
8
radiologic data
8
tracked endoscope
8
distance endoscope
8
endoscope target
8
target structure
8
application error
8
endoscope
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!