Introduction: Open-irrigated radiofrequency ablation catheters with slight differences in tip architecture are widely used, although limited comparative data are available. The purpose of this study was to compare the lesion size and potential complications produced by commercially available open-irrigated catheters in an in vitro porcine heart model.

Methods And Results: Six catheters were tested (Biosense Webster Thermocool, Boston Scientific Open irrigated, St. Jude CoolPath, St. Jude CoolPath Duo, Biosense Webster Thermocool SF, St. Jude Cool Flex) at 20 and 35 W power-control, under 2 different blood flows (0.1 and 0.5 m/s) and at 2 target durations (30 and 60 seconds). A total of 601 lesions were made in 26 in vitro preparations. The tip temperature profile showed significant differences between the catheters (P < 0.001) with the Thermocool SF registering the lowest. Only the surface diameter and the depth at maximum diameter of the lesion were influenced by the design of the ablation electrode. The lesion volume did not show significant differences between catheters for any power, application duration or blood flow condition. Char and pops occurred more often at 35 W with only slight differences between the catheters.

Conclusions: Tip design of the 6 different irrigated catheters does not affect the lesion total volume, although a slight difference in lesion geometry in terms of surface diameter and depth at maximum diameter is present. The catheters show a slight different in vitro safety profile.

Download full-text PDF

Source
http://dx.doi.org/10.1111/jce.12175DOI Listing

Publication Analysis

Top Keywords

open-irrigated radiofrequency
8
radiofrequency ablation
8
catheters slight
8
slight differences
8
biosense webster
8
webster thermocool
8
jude coolpath
8
differences catheters
8
surface diameter
8
diameter depth
8

Similar Publications

Article Synopsis
  • A study examined whether adjusting radiofrequency (RF) energy based on left atrial wall thickness (LAWT) during circumferential pulmonary vein isolation (CPVI) improved outcomes for patients with paroxysmal atrial fibrillation (PAF).
  • 212 patients were divided into two groups: one received LAWT-guided CPVI while the other underwent conventional CPVI, both using specific RF delivery protocols.
  • After an average follow-up of 13.4 months, results showed no significant differences in clinical recurrence rates, major complications, and other procedural outcomes between the two groups, indicating that the tailored approach did not enhance safety or efficacy.
View Article and Find Full Text PDF

Background: At the same conditions of delivered power and contact force, open-irrigated radiofrequency ablation catheters are believed to create deeper lesions, while non-irrigated ones produce shallower lesions. This ex vivo study aims to directly compare the lesion dimensions and characteristics of an irrigated ablation catheter with a flexible tip and a non-irrigated solid-tip catheter.

Methods: Radiofrequency lesions were induced on porcine myocardial slabs using both open-tip irrigated and non-irrigated standard 4 mm catheters at three power settings (20 W, 30 W, and 40 W), maintaining a fixed contact force of 10 gr.

View Article and Find Full Text PDF

The QDOT MICRO™ Catheter is a novel open-irrigated contact force-sensing radiofrequency ablation catheter. It offers very high-power short-duration (vHPSD) ablation with 90 W for 4 s to improve safety and efficacy of catheter ablation procedures. Although the QDOT MICRO™ Catheter was mainly designed for pulmonary vein isolation (PVI) its versatility to treat atrial fibrillation (AF) and other types of arrhythmias was recently evaluated by the FAST and FURIOUS study series and other studies and will be presented in this article.

View Article and Find Full Text PDF

Background: Since the local impedance (LI) of the ablation catheter reflects tissue characteristics, the efficacy of higher power (HP) compared to lower power (LP) in LI-guided ablation may differ from other index-guided ablations.

Objective: This study aimed to assess the efficacy of HP ablation in LI-guided ablation of atrial fibrillation (AF).

Methods: A prospective observational study was conducted, enrolling patients undergoing de novo ablation for AF.

View Article and Find Full Text PDF

Ambient circulation surrounding an ablation catheter tip affects ablation lesion characteristics.

J Cardiovasc Electrophysiol

April 2023

Section of Cardiac Electrophysiology, Division of Cardiovascular Medicine, Stanford University, Stanford, California, USA.

Introduction: The association between ambient circulating environments (CEs) and ablation lesions has been largely underexplored.

Methods: Viable bovine myocardium was placed in a saline bath in an ex vivo endocardial model. Radiofrequency (RF) ablation was performed using three different ablation catheters: 3.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!