Objective: To assess a commercially available point-of-care assay for measurement of bovine cardiac troponin I (cTnI) concentration in blood and plasma samples.

Sample: Prepared bovine plasma standard samples with known concentrations (0 to 1.0 ng/mL) of cTnI and blood and plasma samples obtained from 28 healthy 2.5-month-old Holstein calves.

Procedures: Coefficients of variation were calculated for concentrations of cTnI in prepared standards determined with the point-of-care assay, and values were compared with the known concentrations. The cTnI concentrations in blood samples obtained from calves determined with the point-of-care assay were compared with cTnI concentrations in plasma samples obtained from those animals determined with a validated immunoassay.

Results: The coefficients of variation of cTnI concentrations determined for prepared standards by use of the point-of-care assay were low (< 20%) for standards with cTnI concentrations ≥ 0.025 ng/mL. The blood cTnI concentrations determined with the point-of-care assay were not significantly different from the plasma cTnI concentrations determined with the validated immunoassay.

Conclusions And Clinical Relevance: Results of this study indicated the point-of-care assay had high precision for determination of cTnI concentrations in most evaluated prepared bovine plasma standard samples. The point-of-care assay may be useful for determination of circulating concentrations of cTnI in cattle.

Download full-text PDF

Source
http://dx.doi.org/10.2460/ajvr.74.6.870DOI Listing

Publication Analysis

Top Keywords

point-of-care assay
32
ctni concentrations
28
concentrations ctni
12
determined point-of-care
12
concentrations determined
12
ctni
11
concentrations
11
point-of-care
8
commercially point-of-care
8
assay
8

Similar Publications

Point-of-care ultrasound in the diagnosis of hepatic gas gangrene.

J Ultrasound

January 2025

Argentinian Critical Care Ultrasonography Association (ASARUC), Buenos Aires, Argentina.

Hepatic gas gangrene (HGG) is a rare but life-threatening condition typically caused by anaerobic bacteria such as Clostridium perfringens, though Gram-negative bacteria like Escherichia coli and Klebsiella species have also been implicated. Traditionally diagnosed via computed tomography (CT), point-of-care ultrasound (POCUS) has emerged as a valuable tool in critical care settings for its non-invasive, bedside utility. We report the case of a 51-year-old female with choledochal syndrome secondary to cholangiocarcinoma who developed HGG following left extended hepatectomy and biliary reconstruction.

View Article and Find Full Text PDF

Introduction: To target psychological support to cancer patients most in need of support, screening for psychological distress has been advocated and, in some settings, also implemented. Still, no prior studies have examined the appropriate 'dosage' and whether screening for distress before cancer treatment may be sufficient or if further screenings during treatment are necessary. We examined the development in symptom trajectories for breast cancer patients with low distress before surgery and explored potential risk factors for developing burdensome symptoms at a later point in time.

View Article and Find Full Text PDF

Background: Point-of-care ultrasound (POCUS) can be used in a variety of clinical settings and is a safe and powerful tool for ultrasound-trained healthcare providers, such as physicians and nurses; however, the effectiveness of ultrasound education for nursing students remains unclear. This prospective cohort study aimed to examine the sustained educational impact of bladder ultrasound simulation among nursing students.

Methods: To determine whether bladder POCUS simulation exercises sustainably improve the clinical proficiency regarding ultrasound examinations among nursing students, evaluations were conducted before and after the exercise and were compared with those after the 1-month follow-up exercise.

View Article and Find Full Text PDF

Quantitative and Visual Benefits of Data-Driven Motion Correction on Oncologic PET/CT: A Prospective Cross-sectional Study.

Acad Radiol

January 2025

Department of Radiology, Washington University School of Medicine, 510 S. Kingshighway Blvd, St. Louis, MO 63110 (S.I., M.A.T., M.I., C.S., R.L., A.H., R.L.W., T.J.F.). Electronic address:

Rationale And Objective: Conventional positron emission tomography (PET) respiratory gating utilizes a fraction of acquired PET counts (i.e., optimal gate [OG]), whereas elastic motion correction with deblurring (EMCD) utilizes all PET counts to reconstruct motion-corrected images without increasing image noise.

View Article and Find Full Text PDF

Background: Psoriasis is a chronic disease with a prevalence of 3% in the general population. The high prevalence of psoriasis has prompted the study of its comorbidities in recent decades. However, no studies have ever analyzed comorbidity patterns including all chronic diseases in psoriatic patients.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!